17
Version1.0 rev 18 Feb 2015
Rotr
SECTION 7. SPECIFICATIONS
7.1. P
OWER
S
UPPLY
This product must be powered via the rear 3-pin IEC C14 connector from an a.c. mains power supply with
the following ratings:
Parameter
Specification
Voltage Range
24 V DC
Frequency Range
50-60 Hz
Power Rating (max.)
120W
* Note that this is a technical term to say that the product will expect to be plugged into a typical mains
socket and experience the typical transient voltages that appear there.
7.2. O
PERATING
C
ONDITIONS
Parameter
Specification
Location
Indoor use
Operating Temperature
23°C ± 5°C
Storage Temperature
0-50
o
C
Operating Relative
Humidity
70% non-condensing
Altitude
Up to 2000 m
Pollution Degree
2*
* Note that this is a technical term to say that normally only non-conductive pollution occurs; occasionally,
however, a temporary conductivity caused by condensation must be expected.
•
This product has not been designed to operate correctly in direct sunlight, so it should be avoided.
•
The term “indoor use” also implies that it is not designed to expect water spills.
•
Whilst running an experiment try to keep room temperature as stable as possible.