background image

Unpack the Spectrograph

2

3

1

Prepare your Facility

Install Software (SDK2 or Solis)

Shamrock 500i & 750 Quick Start Guide

Spectrograph

Accessories

Shamrock 500i or 

750 Spectrograph 

(model as ordered)

Power Supply  USB 2.0 Cable 

Type A to Type 

B (3 m)

Camera(s)

(when applicable)

Documentation

scan or visit:

andor.oxinst.com/

welcome

 

Software

(copy supplied)

Hex Key Set

 

You must have administrator access on your PC to perform this installation.

These instructions cover installation of both Andor Solis and Andor SDK if ordered. A copy of the software is provided 

within the spectrograph box. Additionally, the latest software files are available to download on request, please sign 

in or create an account at 

andor.oxinst.com/downloads

Windows: Switch on the PC, run the setup exe from the supplied copy or the downloaded file: e.g. AndorSDK2Setup-

<version>.exe

1. 

Confirm the version of software (minimum is v2.1.x).

2. 

Follow the on screen prompts.

3. Select the installation directory when prompted.

4. 

If prompted for select the required hardware (spectrograph and/or camera) type.

5. 

On the final window click on “Install”. Continue to follow the on screen prompts.

Shamrock will require SDK2 or Solis installation, please see quick start guide for specific 

camera model software requirements.
Online access to technical documentation is available at: 

andor.oxinst.com/downloads

Please follow these guidelines when unpacking the Shamrock: 

1. 

The Shamrock should be placed in the environment it will be used in for a few hours before 
adjustment and first use. 

2. 

Environment should be dust free. 

3. 

Spectrograph should be placed on a flat, stable support (e.g. optical table); Shamrock feet 

are designed to be secured on an optical table with appropriate screws.

The Shamrock 500i or 750 can be supplied with a range of accessories as specified at the time of ordering. Please 
check the supplied items against the packing list to ensure all items are present.

Handle the spectrograph and accessories 
with care.

Use correct power supply and cables.

Observe anti-static precautions when 

installing the spectrograph, camera and 
accessories.

Operate at a constant temperature between 

0°C and 30°C (temperature fluctuation during 
operation may result in calibration drift).
Store between -25°C and 50°C.

Ensure sufficient clearance around 

spectrograph and power supply vents 

(Do 

not

 block vents when operating on 

benchtop).

Please observe the following warning when 
lifting the spectrograph:

Operation of the spectrograph close 

to intense pulsed sources (e.g. plasma 
and X-ray sources) may compromise 

performance.

Keep dry, <70% humidity non-condensing.

Refer to the user manual for further information