Microstat He Manual
trained personnel, when handling heavy system components.
Appropriate lifting equipment and personal protective equipment (PPE) must be provided at the installation site for the duration of the
system installation.
3.10
Magnetic fields
Powerful magnetic fields can interfere with the operation of cardiac pacemakers. Do not approach the equipment if
you wear a cardiac pacemaker.
Systems with the magnet option contain powerful electromagnets. These electromagnets can produce a powerful magnetic field, which
can interfere with the operation of cardiac pacemakers in the vicinity of the system, causing death or serious injury.
Risk to personnel with cardiac pacemakers.
Where necessary, the appropriate warning signs should be in place around the installation site. Personnel who have a cardiac pacemaker
must not approach the system at any time, even when it is powered off. The magnet can exert a force on nearby ferromagnetic objects,
which presents a hazard to personnel if these objects are free to move.
3.11
Asphyxiation
Helium and nitrogen can displace the oxygen from air and cause death by asphyxiation. Ensure that adequate
ventilation is provided.
Areas where these chemicals are stored or used must be well ventilated to avoid the danger of suffocation. Oxygen level detection
equipment should be installed in suitable locations to warn personnel if the oxygen concentration falls below a threshold value. Take
precautions to prevent spillage of liquid cryogens.