Operator's Manual
Issue 14
/
Mar 2016 / UMC0071
Mercury
iPS
©2016 Oxford Instruments NanoScience. All rights reserved.
Page
156
17 TECHNICAL SPECIFICATIONS
This chapter lists the technical specifications of the
Mercury
iPS.
17.1 Physical specification
Item
Specification
Width
426 mm
Depth
481 mm
Height
131 mm
Weight
14.0 kg
17.2 Electrical power supply
Item
Specification
Mains supply voltage
100 VAC to 240 VAC
Mains supply frequency
50 Hz to 60 Hz
Maximum supply power
800 W
Maximum mains voltage fluctuation
±10% of nominal voltage
Compliance Standard
CE,
C
ETL
US
(UL and CSA recognised standard)
Over current protection
Internal – no user changeable fuses inside.
Category II transient overvoltages can typically be present on the mains supply.
17.3 Magnet outputs
Item
Specification
Main output current range
±60.0A nominal ±62.5A peak per Mercury unit (up
to ±600A (10V))
Main output current resolution
< 9µA per Mercury unit
Main output current stability
< ±2mA per Mercury unit
Main output current drift
< 0.1mA/Hr/degC (Temperature controlled sense
resistor and error amplifier)
Main output current accuracy
Error < ±0.13%
Main output current ripple
PWM output control prevents supply ripple
reaching the current output
Typical output current noise (at 60A)
-48.4dBm (0.51mA)