background image

6

5

To enter a weighed tare: 
1.
  Place an empty container on the platter and press the Tare key. The display will return to zero,   

eliminating the weight of the container. The Zero light will go off and the Tare light will be lit.

2.  To clear the weighed tare, remove all weight from the scale. The display will show a negative 

weight. Press the TARE key to return the display to zero, eliminating the weight of the container. 

The Tare light will go off and the Zero light will be lit. 

To manually enter a known tare:
1.
   Use the number keys to input the tare weight. Your entry will be displayed in the “WEIGHT” 

display window. Then press the TARE key to confirm or press the ZERO/ON/OFF key to exit 

and not confirm. Example: With the platter empty, entering 100g and pressing the TARE key will 

display “–0.100” kg.

2.   To recall the previously stored tare weight, press the RC.PLU key. Press the ZERO/ON/OFF key 

to return to weighing.

   Note: The previously stored tare weight can only be recalled with the RC.PLU key when in tare 

setting mode.

3.   To clear the manually entered tare weight, enter “0” and press the TARE key to confirm.
NOTE: For a tare greater than 10kg, the weight window will display “dSP.oVr” when the weight 

is  removed from the platter. The tare function is still working properly, but since the “WEIGHT” 

window cannot display “-10.0000”, “dSP.oVr” indicates that the displayed value is beyond the 

range of display.

ACCUMULATION MODE
The accumulation function allows storage of weighed piece quantities and the summation of those 

quantities. 
1.   With an object on the scale, press and release the ACCU key to add the displayed value to the  

accumulated total.  

2.   The indicator will display “Add” in the WEIGHT window, the PIECE WEIGHT window will display 

the number of accumulation instances, and the COUNT window will display the number of 

accumulated piece quantity. 

3.   Press and hold the ACCU key for 4 seconds to display the accumulated values. The WEIGHT 

window will display “Add”, the PIECE WEIGHT window will display the number of accumulation 

instances, and the COUNT window will display the accumulated quantity.

4.   To clear accumulated quantities, press and hold the 0/CLEAR key while the accumulated 

values are being displayed until they reset.

NOTE: Powering off the scale will also clear the accumulated values.
NOTE: Accumulated quantities or instances over 999999 will display “Err01”.

COUNTING MODE
The counting function calculates and displays the piece quantity of the load that is being weighed.  

You can enter the piece weight by one of two methods:
1.   To enter a piece weight by weighing a known quantity: 
a.   Place a known quantity of objects onto the platter and press the PCWT/SPL key for 4 seconds 

to enter piece weight calculation mode.

b.   The WEIGHT window will display the weight, the PIECE WEIGHT window will display “SPL.PCS”  

(Sample Pieces), and the COUNT window will display the known quantity that you will input.

OXD-844-2040K-INSTRUCTIONS.indd   6

OXD-844-2040K-INSTRUCTIONS.indd   6

29/06/2020   10:07

29/06/2020   10:07

Summary of Contents for OXD-844-2040K

Page 1: ...K cromwell co uk Battery or mains powered Suitable for office Postroom stores applications Large clear digital display Dual reading metric or imperial Capacities 6kg 12lb OXD 844 2040K INSTRUCTIONS in...

Page 2: ...the weighing plate In some circumstances the internal sensor may be damaged by inappropriate use The item to be weighed should be placed carefully onto the scales and the display allowed to stabilise...

Page 3: ...range calibration zero point 15 FS ZERO key range power on zero 5 FS Tare Range Full capacity Stabilisation Time 5 seconds Operating Temperature 32 to 105 F 0 to 40 C Humidity Range 90 relative humidi...

Page 4: ...s for 4 seconds to clear data or accumulated values Enter Confirms the operation or saves the data Print Outputs data via the RS232 port Accu Accumulates the current weighed values Press for 4 seconds...

Page 5: ...scales allow 10 minutes for them to warm up before use for optimum results 2 Press the ZERO ON OFF key to turn on the scales The self check will run and the scales will display a zero reading The sca...

Page 6: ...MULATION MODE The accumulation function allows storage of weighed piece quantities and the summation of those quantities 1 With an object on the scale press and release the ACCU key to add the display...

Page 7: ...tone will sound 1 From normal weighing mode press the Hi Lo key to enter this mode The WEIGHT window will show the current weight and the PIECE WEIGHT window will show Hi PCS High limit pieces The COU...

Page 8: ...o confirm or press the ZERO ON OFF key to exit this mode a If you selected PCT 0 to calculate the piece weight the COUNT window will show SPL SAMPLE Place samples on platter and the PIECE WEIGHT windo...

Page 9: ...reading is stable press PRINT key to output the data via RS232 interface in the following format Business xxxxx Product xxxxx ID xxxxxx Date yy mm dd Time hh mm Gross xxx xxx kg lb Tare xxx xxx kg lb...

Page 10: ...cator stops flashing press the ENTER key to confirm the standard weight point calibration or press the ZERO ON OFF key to exit this mode After the standard weight point is confirmed the scale will aut...

Page 11: ...y range 5 Make sure scale platform is empty Power off scale and power on again Perform zero calibration Err05 Weight is below Power On Zero limit 15 or below ZERO key range 5 Install platform on scale...

Page 12: ...ecycling facility Do not dispose of electrical goods with domestic waste materials as inappropriate disposal may cause potential hazards to the environment and human health For further information ple...