Rev.0 LEPTON
Service
manual
OXYGEN S.p.A.
21
Figure 7/b
Figure 7/c
Figure 7/d
Figure 7/e
REMOVING THE NICKEL-ZINC BATTERIES
1. Remove the cover of the case to access the batteries (2 – Figure 7).
The case is secured by a lock at the front (3 – Figure 7), and a notch at the back.
2. Press down to unhook the metal rod (4 – Figure 7) holding the battery in place (5 –
Figure 7).
3. Using the appropriately-sized tool, disconnect the NEGATIVE pole from the
battery, black cable with the lug on the left (Figure 7/b).
4. Using the same tool, disconnect the POSITIVE pole and the temperature sensor
from the battery red cable on the right (Figure 7/c).
5. Remove the bridges that connect the two battery sets with the same tool (Figure
7/d).
The first battery unit can be moved at this point, holding it by its two sides and