Error in replacing detachable components can lead to unacceptable risks. Replace the
detachable components according to the correct steps in the instructions.
•
Wireless charging will generate heat, and the surface temperature of the charging base and
motor handpiece will rise. It is recommended that the time of contacting motor handpiece
and charging base during wireless charging should not exceed 10 seconds.
•
The device is not intended to be used in areas where liquids are likely to be present at floor
level, such as emergency rooms.
•
Replacement of lithium batteries by inadequately trained personnel or incorrect replacement
could result in a hazard, so please contact local distributors to replace the battery if necessary.
•
The adapter plug can be used to disconnect from the network power supply. Do not position
the device to make it difficult to operate the disconnection device when charging.
1.13 Device safety classification
1.13.1 Type of operation mode: Continuous operating device
1.13.2 Type of protection against electric shock: Class II equipment with internal power supply
1.13.3 Degree of protection against electric shock: B type applied part
1.13.4 Degree of protection against harmful ingress of water: