6.3.1.2 Manual disinfection steps:
1.
Soak the dry, soft cloth with 75% alcohol.
2.
Wipe all surfaces of the motor component, power adapter, charging base and other
components with a wet soft cloth for at least 3 minutes.
3.
Wipe the surface of the component with a dry, soft nap-free cloth.
Note:
•
The cleaning and disinfection must be performed within 10 minutes before use.
•
The disinfectant used must be used immediately, no foaming is allowed.
•
In addition to 75% alcohol, you can use non-residue disinfectants such as Oxytech from
Germany, but you must respect the concentration, temperature and time specified by the
disinfectant manufacturer.
•
After cleaning and disinfecting the motor component, you must install a disposable barrier
before use and repeat steps 1, 2, and 3 to clean the disposable barrier (For detailed
installation steps, see section 2.*).
6.3.2 Post-Op processing
After each use, clean and disinfect the motor component, power adapter, and charging base
within 30 minutes. The specific steps are as follows:
Tools: Nap-free soft cloth, tray
1.
In a clean tray, remove the disposable barrier from the motor component.
2.
Soak the nap-free soft cloth with distilled water or deionized
w
ater, and then wipe all
the surfaces of the components such as the motor component,
power adapter,
charging
base, etc. until the surface of the component is not stained.