Before use, you must carefully read the operating instructions provided by the equipment
manufacturer to familiarize yourself with the disinfection process and precautions.
•
With this equipment, cleaning, disinfection and drying will be carried out together.
•
Cleaning: (c1) The cleaning procedure should be suitable for the product to be treated. The
flushing period should be sufficient (5-10 minutes). Pre-wash for 3 minutes, wash for another 5
minutes, and rinse it twice with each rinse lasting for 1 minute. (c2) In the washing stage, the
water temperature should not exceed
45°C (113°F)
, otherwise the protein will solidify and it is
difficult to remove. (c3) The solution used can be pure water, distilled water, deionized water
or multi-enzyme solution, etc., and only freshly prepared solutions can be used. (c4)
With
the
use of cleaner, the concentration and time provided by manufacturer shall be obeyed. The
used cleaner is neodisher MediZym (Dr. Weigert).
•
Disinfection: (d1) Direct use after disinfection: temperature ≥ 90
°C (194°F)
, time ≥ 5 min or A0
≥3000; Sterilize it after disinfection and use: temperature ≥
90°C (194°F)
, time ≥1 min or A0 ≥
600.
(d2)
For the disinfection here, the temperature is 93
°C (199°F)
, the time is 2.5 min, and
A0>3000.
•
Only distilled or deionized water with a small amount of microorganisms (<10 cfu/ml) can be
used for all rinsing steps. (For example, pure water that is in accordance with the European
Pharmacopoeia or the United States Pharmacopoeia).
•
After cleaning, the chemical residue should be less than 10mg/L.
•
The air used for drying must be filtered by HEPA.
•
Regularly repair and inspect the
disinfector
.