Technical Description Concerning Electromagnetic Immunity Table 2: Guidance & Declaration -
electromagnetic immunity
Guidance & Declaration — electromagnetic immunity
The
model
ProMate™ CL is intended for use in the electromagnetic environment specified below.
The customer or the user of
the
ProMate™ CL should ensure that it is used in such an environment.
Environment.
Immunity test
IEC 60601
Test level
Compliance level Electromagnetic Environment - Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8kV contact ±2,
±4, ±8, ±15kVair
±8kV contact
±2, ±4, ±8,
±15kVair
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
±2kV for power
s
upply lines
±1kV for Input/
Output lines
±2kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC
61000-4-5
±0.5, ±1kV line
to line ±0.5,
±1, ±2kV line
to
earth
±0.5, ±1kV line
to line ±0.5,
±1, ±2kV line to
earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Electrical fast
transient/
burst IEC
61000-4-4