42
Environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic Environment - Guidance
Voltage
dips, short
interruptions
and voltage
variations on
power
supply input
lines IEC
61000-4-11
<5% UT (>95% dip
in UT.) for 0.5 cycle
<5% UT (>95% dip
in UT.) for 1 cycle
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip
in UT) for 250 cycles
Mains power quality should be
that of a typical commercial or hospital
environment. If the user of the model
ProMate™ CL requires continued operation
during power mains interruptions, it is
recommended that the model ProMate™
CL be powered from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30A/m
30A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
NOTE: UT is the A.C. mains voltage prior to application of the test level.
Table 2: Guidance & Declaration - electromagnetic immunity
-
Cont
inued
<5% UT (>95% dip in
UT.) for 0.5 cycle
<5% UT (>95% dip in
UT.) for 1 cycle
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 250 cycles