4.4 Safety Test
!
The emergency stop switch on the NANO polisher controls the operation of the FEMTO 1500
power head when operated in the AUTO mode. Test the emergency stop switch by running
the FEMTO 1500 head in the AUTO mode through the NANO polisher. The following safety
check is considered important:
Test:
Activate main power switch.
Start NANO 1000T/ FEMTO 1500
machine in AUTO mode.
Depress emergency stop switch.
Proper
Response:
NANO 1000T and FEMTO 1500 head
stops.
Malfunction:
NANO 1000T or FEMTO 1500 polisher
does not lose power.
Corrective
measure:
If system does not power down,
disconnect power supply cord and call
service technician.
Emergency stop switch