- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ▲
I
NSTRUCTION
M
ANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1 520-882-6598 Fax +1 520-882-6599 email: pace@metallographic.com Web: http://www.metallographic.com
24
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
NANO 1000T / FEMTO 1500
Polishing Head
Preset
Speeds
5.6 SEMI-AUTO NANO-FEMTO Polishing Mode
1. Install working wheel and attach grinding papers /
polishing cloths.
2. Power on the FEMTO and NANO machines
using the power switch located on the back of the
units.
Find and press the AUTO button on the FEMTO
control panel
Set an approximate pressure and head speed on
the FEMTO head.
Load samples.
3. Start the NANO head by using preset speeds
(SPL, SPM, SPH) and pressing RUN/STOP
button. Note speed can be adjusted with up and
down arrow keys.
Adjust the FEMTO speed and force.
The FEMTO head and NANO polisher will start
automatically in this mode, however, it will run
until the RUN/STOP button is pressed.
4. Position flexible water spout over working wheel.
During sample preparation adjust water flow by
turning water control knob as required. Note:
Initial operation of water valve may contain air in
the lines. Turn water on slowly to purge air from
system.
5. Press the RUN/STOP button on the NANO
controller to stop the polishing operation.
Set to
Auto
mode
Adjust
head
speed
Adjust
polishing
force
Start/Stop
NANO
manually