-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
▲
I
NSTRUCTION
M
ANUAL
MEGA
-
T400S
Abrasive Cutter
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1
-
520
-
882
-
6598 Fax +1
-
520
-
882
-
6599 email: pace@metallographic.com Web: https://www.metallographic.com
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
6.1 Trouble Shooting Error Codes
20
Error Code
Code Description
Solution
ERR 01
IPM error
Disassemble and return to supplier
ERR 02
Over current
Check the setting parameters
ERR 03
EEPROM error
Rewrite the default value
ERR 04
Current sampling
Rewrite parameters
ERR 05
(GF)
Bus under or over volt-
age
Check the voltage and parameters
ERR 06
Communication anomaly
Check the wire connection
ERR 07
Locate failed
Check the encoder signal; Or the current feedback
wrong. Reposition and try again.
ERR 08
Lost step when running
Normally, the reverse angle is 2 degrees, and the re-
verse angle is 20 degrees when positioning will be
wrong. Reposition and try again.
ERR 09
Overload
Cutting too fast, too much load, or reduce the speed.
ERR 10
Module temperature is
too high
Check whether the fan works properly
ERR 12
Power failure
Disassemble and return to supplier
ERR 13
Encode failure
Check wire connection, or the encoder was damaged
ERR 14
Motor get stuck
Check the motor or cutting speed is too high