Calibration
°F/°C READOUT DEFAULT
6.
Press and release the TIP SET Key. The Digital Readout will display "S - X" (X = "-" or 1-9). Either the
°F or °C LED will be on. This is the default temperature scale of the Digital Readout (e.g., if the °C LED
is on, the Digital Readout will display Tip Temperatures and Tip Offset Constant values in °C).
7.
Press and release the °F/°C Key to change the default.
Each subsequent press and release of the key will change
the default.
AUTOMATIC SETBACK
Figure 20. Change Temperature Default
8.
As received from the factory, "S - -" will be displayed indicating that the Automatic Temperature
Setback is turned off. A "1" thru "9" appearing on the right side of the Digital Readout indicates time to
Automatic Setback in increments of 10 minutes. For example, "S-3" would indicate that any Active
Channel will set back its handpiece’s Set Tip Temperature to 177°C (350°F) after 30 minutes of
handpiece inactivity (non-use). To change the time period or turn the Automatic Temperature Setback
feature off or on, use the Scroll Keys. Press the Scroll Up Key to increase the time period and/or enable
the feature. Press the Scroll Down Key to decrease the time period or disable the feature.
9.
Press the TIP SET Key to store the °F/°C default and Automatic Temperature Setback time value in
system memory. The Digital Readout will revert to "CAL" and only the CH 1 LED will remain lit.
38