GENERAL INFORMATION
SPECIFICATIONS
POWER REQUIREMENTS
MBT 250
-
Version operates on 97-127 VAC, 50/60 Hz.
185 Watts, 1.6 Amp max, 100% Duty Cycle, Motor on.
MBT250E
-
Version operates on 196-253 VAC, 50 Hz.
212 Watts, 0.92 Amp max, 100% Duty Cycle, Motor on.
PHYSICAL PARAMETERS
Size:
13.5 cm H x 16.5 cm W x 26 cm D (5.3"H x 6.5"W x 10.25"D)
Weight:
4.5 Kg. (10 Lbs.)
VACUUM AND AIR
(motor operated systems)
Measurements at front panel
AUTO SNAP-VAC
and
CONTROLLABLE PRESSURE
Ports.
Vacuum Rise Time:
Evacuates 33 cc (2 cubic inch) volume
to 25 cm Hg. (10 in. Hg.) in 150 ms.
Vacuum:
51 cm Hg. (20 in. Hg.) (nominal)
Pressure:
.48 Bar (7 P.S.I.) (nominal MAXIMUM setting)
AirFlow:
9 SLPM (0.32 SCFM)MAXIMUM
TEMPERATURE SPECIFICATIONS
Tip Temperature Range:
38°C to 482°C (100°F to 900°F) nominal (see note).
Digital Readout Resolution:
±1° (°C or °F)
Tip Temperature Stability:
±1.1°C (2°F) at Idle from Set Tip Temperature.
5