6.6.11
PULSE Parameters
Pulse transients can be used to create repetitive events for endurance testing of AC and DC
powered products.
The following parameters are available in a PULSE VALUES screen.
PARAMETER
Range
Unit
Description
Frequency
15 ~ 1200
Hz
Frequency
Voltage AC
0 ~ 300
Vrms
AC Voltage
Waveform
1 ~ 200
Waveform number
Phase
0.0 ~ 359.9
Deg
Phase angle for phases B & C
Table 6-17: Available STEP Transient Parameters
The following soft keys are available on the PULSE VALUES screen.
SOFT KEY
Description
Use Setpoint (SK1)
Starts the transient from the “Run from step#” row number
Run Screen(SK2)
Displays the PULSE EXECUTION screen.
Link Phases (SK3)
Toggles between phase A, B, C or Coupled ABC
Example (SK4)
Displays a graphical representation of the STEP parameters
Back (SK5)
Returns to previous screen
Table 6-18: Available STEP PROGRAM screen soft keys