LMX SERIES™ OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
Page 354 of 463
Command Syntax
PROGram:TRANsient:TRIGger:INput:AUTOrun <b>
Description
When the trigger input is active and segments are running state, each
trigger event (pulse) will automatically start a new sequence, without
the need of a new RUN command before each trigger. The RUN
command has to be executed only once, and then each subsequence
trigger input event will cause a new segment sequence to run.
Note:
the trigger signal is level-sensitive; hence if it is kept high,
it will continuously issue a trigger.
Parameters
[ 0 | OFF | 1 | ON ]
Parameter Format
<b>
Example
PROG:TRAN:TRIG:IN:AUTO 1
Query Format
PROGram:TRANsient:TRIGger:INput:AUTOrun?
Returned Data Format
<b>
Query Example
PROG:TRAN:TRIG:IN:AUTO?
1
Command Syntax
PROGram:TRANsient:TRIGger:OUTput
Description
This command causes a trigger output pulse to be generated when a
transient execution is started. Note that this output is used as a
function strobe during steady state operation.
Parameters
[ 0 | OFF | 1 | ON ]
Parameter Format
<b>
Example
PROG:TRAN:TRIG:OUT 1
Query Format
PROGram:TRANsient:TRIGger:OUTput?
Returned Data Format
<b>
Query Example
PROG:TRAN:TRIG:OUT?
1