11.4
Calibration Procedures
Calibration of output and measurements is performed by a single procedure so there is no need
to calibrate each separately. Calibration involves voltage and current full scale and offset on
each phase. For three-phase mode, this means there are twelve calibration points, for single-
phase mode, there are four.
The calibration can be performed manually using the Calibration menu (refer to Section 6.8.9 for
the location of the Calibration menu) or through the build-in web server using the web browser
interface. Refer to Section 10.10.10, “Calibration” for access to the Web based calibration
screens.
11.4.1
Voltage Calibration - Offset
AC Line
Input
DMM (optional)
A B C N N N
Short A, B & C to N for
Voltage Offset
Calibration
Table 11-2: Setup for Voltage Offset Calibration
The procedure to calibrate voltage offset is:
1.
Select the Calibration entry from the “
System
” menu
2.
Press the “Calibrate” soft key to enter calibration mode