2
Contents
Safety and Information Symbols ...........................................................................................3
1. Safety Instructions .............................................................................................................3
1.1 Personal Safety ..............................................................................................................3
1.2 Safe Use of the Palltronic Flow Check II Unit..................................................................3
2. Introduction ........................................................................................................................4
2.1 Key Features and Benefits ............................................................................................4
2.2 Standard Functions ........................................................................................................4
3. Using the Palltronic Flow Check II Unit ............................................................................5
3.1 Features ........................................................................................................................5
3.3 Display Function ............................................................................................................6
3.3 Test Preparation ............................................................................................................6
3.4 Performing a ‘Flow Check’ Test......................................................................................7
3.5 Understanding the Test Results......................................................................................7
4. Maintenance .......................................................................................................................8
4.1 Changing the Battery ....................................................................................................8
5. Troubleshooting..................................................................................................................8
5.1 Preliminary Troubleshooting............................................................................................8
5.2 Continued Difficulties......................................................................................................9
Appendix A – Atmospheric Pressure Correction Table ......................................................10
Page 1: ...USD 2642 Instructions For Use Palltronic Flow Check II Unit Part numbers FC02 FC02M and FC02H...
Page 2: ...its 4 2 2 Standard Functions 4 3 Using the Palltronic Flow Check II Unit 5 3 1 Features 5 3 3 Display Function 6 3 3 Test Preparation 6 3 4 Performing a Flow Check Test 7 3 5 Understanding the Test Re...
Page 3: ...USD2363 Please contact your local Pall office or distributor if you require a copy of the latest revision of any of these documents 1 1 Personal Safety Appropriate protective clothing should be worn...
Page 4: ...Within the pharmaceutical industry there is a regulatory requirement to integrity test filters both before and after use in a production environment Therefore a fully functional integrity test instrum...
Page 5: ...l attached to the enclosure displays the certified flow rate using air at an atmospheric pressure of 1000 mbar 14 5 psi the associated test pressure and the unit serial number The display is switched...
Page 6: ...to extend the battery life The unit displays the pressure in mbar only The test pressure is normally 2000 mbar 29 psi 200 kPa 2 039 kp cm2 3 3 Test Preparation The integrity test assembly comprising...
Page 7: ...ass fail result at the end of the test should be ignored refer instead to Section 3 5 for the interpretation of the result Select a Flow Check test on the Palltronic integrity test instrument for more...
Page 8: ...failed message can be erased by pressing the Start button for more than 5 seconds to switch the unit off Used batteries should be discarded in accordance with relevant local and national regulations...
Page 9: ...for the ambient pressure Check that the test parameters especially test pressure are set correctly these details are available with the Palltronic Flow Check II unit see Section 3 4 for more informat...
Page 10: ...700 107 0 106 0 105 0 104 0 103 0 102 0 101 0 100 0 99 0 98 0 97 0 96 0 95 0 94 0 93 0 92 0 91 0 90 0 89 0 88 0 87 0 86 0 85 0 84 0 83 0 82 0 81 0 80 0 79 0 78 0 77 0 76 0 75 0 74 0 73 0 72 0 71 0 70...
Page 11: ...ain Sweden Switzerland Taiwan Thailand the United Kingdom the United States and Venezuela Distributors in all major industrial areas of the world The information provided in this literature was review...