Safety and Information Symbols

Safety and information is identified in this instruction manual with the following symbols:

Warning! Identifies a dangerous situation where there is a safety risk and personal injury may follow.

Important: Identifies a potentially dangerous situation, a risk to the equipment or other important
information.

Note: Identifies user tips and other useful information.

1.

Safety Instructions

Warning! Please read the following instructions carefully and completely before using the
Palltronic Flow Check II unit.

Please also read the Operating Instructions for the applicable integrity test instrument which will
be used with the Palltronic Flow Check II unit to ensure safe use. The list of literature is as follows:

Table 1

Literature

Document Title

Number

Palltronic Flowstar IV Integrity Test Instrument: Instructions for Use

USD2594

Palltronic Flowstar XC Integrity Test Instrument: Operating Instructions

USD2005

Palltronic Flowmod Filter Integrity Test Instrument: Operating Instructions

USD2223

Palltronic Flowstar MUX Integrity Test System: Operating Instructions

USD2292

Palltronic AquaWIT XC Automatic Integrity Test System: Operating Instructions

USD2363

Please contact your local Pall office or distributor if you require a copy of the latest revision of
any of these documents.

1.1

Personal Safety

• Appropriate protective clothing should be worn by operators (e.g., gloves and safety goggles)

when working with chemicals and compressed gas supplies during filter integrity testing.

• All pneumatic tubing used for integrity testing and with the Palltronic Flow Check II unit should

be rated to withstand the gas pressure used (up to 8 bar or 116 psi). 

• The operator should check the tubing and connectors for damage before each use.

1.2

Safe Use of the Palltronic Flow Check II Unit

• The Palltronic integrity test instrument uses compressed air when in use and, therefore, caution

is necessary when disconnecting pneumatic connections from the Palltronic Flow Check II unit.
It is the user’s responsibility to ensure that an oil-free regulated gas supply is used during
operation of the Palltronic integrity test instrument and Palltronic Flow Check II unit.

• The maximum operating pressure of the Palltronic Flow Check II unit is 3.0 bar (43.5 psi) at a

maximum operating temperature of 40 °C. 

• Do not exceed the maximum operating pressure.

• Compressed gas exits the Palltronic Flow Check II unit at low velocity through the ‘VENT’

port. To ensure correct and safe operation, the ‘VENT’ port must be open to atmosphere.

www.pall.com/biopharm

3

Summary of Contents for Palltronic Flow Check II

Page 1: ...USD 2642 Instructions For Use Palltronic Flow Check II Unit Part numbers FC02 FC02M and FC02H...

Page 2: ...its 4 2 2 Standard Functions 4 3 Using the Palltronic Flow Check II Unit 5 3 1 Features 5 3 3 Display Function 6 3 3 Test Preparation 6 3 4 Performing a Flow Check Test 7 3 5 Understanding the Test Re...

Page 3: ...USD2363 Please contact your local Pall office or distributor if you require a copy of the latest revision of any of these documents 1 1 Personal Safety Appropriate protective clothing should be worn...

Page 4: ...Within the pharmaceutical industry there is a regulatory requirement to integrity test filters both before and after use in a production environment Therefore a fully functional integrity test instrum...

Page 5: ...l attached to the enclosure displays the certified flow rate using air at an atmospheric pressure of 1000 mbar 14 5 psi the associated test pressure and the unit serial number The display is switched...

Page 6: ...to extend the battery life The unit displays the pressure in mbar only The test pressure is normally 2000 mbar 29 psi 200 kPa 2 039 kp cm2 3 3 Test Preparation The integrity test assembly comprising...

Page 7: ...ass fail result at the end of the test should be ignored refer instead to Section 3 5 for the interpretation of the result Select a Flow Check test on the Palltronic integrity test instrument for more...

Page 8: ...failed message can be erased by pressing the Start button for more than 5 seconds to switch the unit off Used batteries should be discarded in accordance with relevant local and national regulations...

Page 9: ...for the ambient pressure Check that the test parameters especially test pressure are set correctly these details are available with the Palltronic Flow Check II unit see Section 3 4 for more informat...