20
l
The margin of error of the Calibrated
YSI Glucose Analyser measurement is
0.2 mmol/L when the blood glucose
concentration is higher than 5.6 mmol/L,
or 0.0016 mmol/L when the blood glucose
concentration is lower than 5.6mmol/L.
l
Test results produced on capillary whole
blood samples by the PalmDoc I system
are compared with the results of the
corresponding plasma samples tested by
the calibrated YSI 2300 Glucose Analyser.
Storage and handling
Please take the following precautions to
ensure your PalmDoc blood glucose test
strips are effective.
l
Prior to first use, ensure that the vial is
undamaged and closed.
l
Keep the test strip vial out of sunlight
and in a cool, dry place between 4 ~ 40°C
(39 ~104°F). Do not freeze it.
l
Store test strips in their original vial only. Do
not put the test strips in any other container.
l
Close the vial cap tightly immediately
after removing a test strip. This keeps the
remaining test strips fully functional right
up to the expiry date.
03.
Blood glucose test strips usage