background image

 !"

 

)

6.0  Advanced Usage 

 

The ZM-30 antenna analyzer is an instrument that can be used for a number of uses well beyond simple SWR and 

impedance measurements. The following list presents some of these advanced uses.  

 

Tuned Circuit Resonance 

 

Resonance of a tuned circuit can be measured in several ways. 

 

The most accurate method is to connect it as series resonant circuit by 

 putting it in series with a non-inductive (carbon composition or film)  

¼ or ½ watt 51 ohm resistor as shown in Figure 1.  

 

Now connect this to J3 on the Analyzer and tune for lowest SWR.   

The frequency at which this is seen is the tuned circuit resonant frequency. 

 

2. If the individual L and C are connected as a parallel resonant circuit  

and cannot be connected is series as above (for example an antenna trap) 

 the parallel resonance can be estimated.   Connect the parallel resonant 

 circuit as shown in Figure 2. 

 

Connect the circuit to J3 on the Analyzer and again tune for lowest SWR.  

 The measured frequency is slightly lower than the tuned circuit resonant 

 frequency due to stray capacitance of the test setup. 

 

Inductor Q 

 

The Q of an RF inductor can be measured with a very simple setup.  

First measure the inductive reactance XL of the inductor and record this 

value.  Now connect it to the Analyzer as shown in Figure 3.  Capacitor C 

must be chosen to resonate with L at the frequency where you want to  

measure the inductor’s Q.  The Inductor and Capacitor Measurement section  

of this manual shows how this capacitor value can be determined. 

 

Now tune the Analyzer for the lowest R (resistance) value with a reading of  

zero X (reactance).  If R is above 10 ohms you can now calculate inductor Q 

using the formula: 

 

 

 

 

If R is less than 10 ohms a slightly different method needs to be used.  In this 

case use the test setup shown in Figure 1.  Adding the non-inductive (carbon  

compostion or film) ¼ or ½ watt 51ohm resistor allows more accurate measurement  

of the series resistance of the inductor. 

 

Again tune the analyzer for lowest R (resistance) value with a reading of  zero  

X (reactance).  Next subtract 51 ohms from the measured R value and use 

This new resistance in the above formula to calculate the Q value. 

 

Transmission line Stub Lengths  

 

Measurement of quarter and half wave transmission line stubs can be performed regardless of the transmission line 

characteristic impedance.  The method relies on the fact that an open-circuited quarter wavelength line or a short-

circuited line acts like a precise short circuit at the chosen frequency of operation. 

 

With either type of feedline first cut the feedline to about 10% longer than the desired length, taking the appropriate 

velocity factor into account.  The velocity factor of common feedlines is available from manufacturer’s literature or 

51

 

TO 

J3 

Figure 1 

51

 

TO 

J3 

Figure 2 

TO 

J3 

Figure 3 

Q =  XL 

Summary of Contents for ZM-30

Page 1: ...0 Overview 2 2 0 Features 3 3 0 Connections Controls 4 4 0 Operation 5 5 0 Basic Usage 9 6 0 Advanced Usage 10 7 0 Technical Support 11 Appendix A Specifications 12 Appendix B Loading New Software 13 Appendix C Calibration 15 ...

Page 2: ...icrocontroller retains the frequency and corresponding results throughout the measurement period and at the end of the scan the microcontroller statically displays the SWR and impedance at the resonant frequencies Thus with a press of a button the antenna analyzer is able to quickly and automatically determine the frequencies for which the antenna system is best matched and display the associated ...

Page 3: ...x impedance at selected frequencies Audible tone indicates resonance lower tone indicates lower SWR May be enabled disabled Serial port connects to PC for field upgradable software Battery operated for convenient field use or can use external 9 16V for bench operation Other uses of the Instrument Measure antenna bandwidth Measure feedpoint impedance Measure ground loss Measure coaxial cable Loss A...

Page 4: ...t of user interface used to select menu items and frequency settings Audible Tones When enabled tones indicate value of SWR MODE User can change the ZM30 to any of the 4 available operating modes Impedance default Capacitance Inductance VFO SETUP Provides menu of user configurable settings and functions Custom Scan settings Audio tone Software Load Calibrate Debug Mode Power Off Time Backlight con...

Page 5: ... user interface consists of four pushbuttons one for each instrument function as described in the next section and a rotary dial that is used to select various options presented in a menu like manner You use the function pushbuttons and the dial to operate the ZM 30 just press a function button and rotate the dial and the various menu options for that function are sequentially displayed on the LCD...

Page 6: ...the dial pushbutton while simultaneously turning the dial The cursor will move to any of the seven available digits allowing subsequent up down adjustment of that digit with the dial pushbutton is released Rotating the dial clockwise will increase the digit value and correspondingly the signal generated by the unit Counter clockwise rotation will decrease the digit and the generated signal When th...

Page 7: ...en the inductance display screen is shown as illustrated below when a 2 2 uH capacitor is connected to the RF output jack 10 000 00 L 2 2 uH The top line indicates the frequency The second line is the computed value for the load connected to the output jack Inductance values that may be displayed must fall within the measurement range of the instrument Keeping in mind the 600 ohm maximum resistanc...

Page 8: ... the display backlight Default setting Lite Off Time This function prompts the user to enter the number of minutes of inactivity before the display backlight is automatically turned off Default value is 10 minutes Power Off Time This function prompts the user to enter the number of minutes of inactivity before the entire instrument is automatically turned off When the user touches any of the contr...

Page 9: ...loss or loading coil loss For example a wave vertical will have a resistance of about 36 ohms at resonance Any higher reading indicates ground loss Similarly shorter antennas when resonated will have lower resistance values Reading a good SWR may mean excess loss and measuring the actual impedance allows gauging just how much loss Adjust antenna tuners The analyzer can be used to adjust an antenna...

Page 10: ...to resonate with L at the frequency where you want to measure the inductor s Q The Inductor and Capacitor Measurement section of this manual shows how this capacitor value can be determined Now tune the Analyzer for the lowest R resistance value with a reading of zero X reactance If R is above 10 ohms you can now calculate inductor Q using the formula If R is less than 10 ohms a slightly different...

Page 11: ...h stub the length is Where L is the length in inches VF is the velocity factor and F is the operating frequency in MHz for the stub Similarly for a quarter wave stub use the formula 7 0 Technical Support If you experience any problems with the ZM 30 or if you have any questions relative to its operation or specifications please contact Palstar at 937 773 6255 Be sure to visit the ZM 30 product web...

Page 12: ...de Band Scan Setup On Off Rotary Encoder Tuning 24 position rotation smooth detentless operation LED Busy 2 line x 16 character LCD display with optional backlight under software control configured in SETUP menus Backlight may be configured to turn off after user set period of inactivity If enabled backlight turns on again with any of the controls are touched Power Control Power applied by press a...

Page 13: ...l modem cable that swaps pins 2 and 3 from end to end 3 Start up the TeraTerm program on your PC If not already set up configure the communications as 9600 8N1 no flow control in the Setup Serial Port menu and be certain to have the line delay set to at least 40ms line Make sure that you have TeraTerm configured to be using the active serial port in your PC Be certain that no other program has con...

Page 14: ...by its respective shunt 5 Connect your ZM 30 to the PC using a standard straight through male female DB9 type serial cable Do not use a null modem cable that swaps pins 2 and 3 from end to end 6 If not already running make sure the TeraTerm program is running on your PC 7 Power up the ZM 30 again by plugging in the external power cable or by inserting the battery that was previously removed 8 You ...

Page 15: ...ircuit on the RF output jack and select yes with the dial 7 See 270 ohm load on displayed on the LCD Place the supplied 274 ohm resistive load on the RF output connector and confirm by selecting yes 8 See Make SWR 5 4 displayed on the first line of the LCD and a number on the second line Adjust the dial to make the number on the second line be as close as possible to 5 4 Press the dial when this n...

Reviews: