Aurora H2O User’s Manual
5
Chapter 2. Installation
2.7 Optional Insertion Probe/Regulator
2.7.1
Description
For natural gas applications that may have entrained particulate and liquid contaminants (especially TEG
carry-over from TEG dryers), BH recommends the use of a
Insertion Probe/Regulator
(see
Figure 9
below) to serve as
the first part of particulate and condensate filtration where the gas is sampled off the pipe. The device combines the
features of a sample tap, a membrane filter and an integral pressure regulator (0-500psig outlet), adjustable at the
sample take-off point. The sample tap housing includes a foot-valve so that, once installed, the assembly can be
removed from a line under pressure for membrane filter replacement as needed.
Figure 9: Insertion Probe/Regulator
Note:
A 29/32” (0.907 inch, 23.1 mm) minimum bore-through clearance is needed on the customer nozzle for
insertion of a 3/4” NPT insertion membrane liquid separator/pressure regulator.
2.7.2 Installation
Note that the
Optional Insertion Probe/Regulator
can be installed only on unpressurized lines. There is no hot-tap
installation for this device.
This procedure is broken down into two steps:
• Installing the Housing
• Installing the Insertion Probe/Regulator Assembly
2.7.2.1
Installing the Housing
The housing, as shown in
Figure 10
below, is supplied with a
Locking Mechanism
to ensure that housing can only be
removed intentionally.
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......