Model CTF878 Startup Guide
23
Chapter 2.
2.1.10 Installing the PDJ Pipe Damping Jacket
If the pipe temperature is over 150°F, you must use a pipe damping jacket with preapplied DMP-3. The jacket is
available in standard pipe sizes (see Table 2-2 and Figure 11 below) or in other sizes on special order.
Figure 11: PDJ Pipe Damping Jacket (Dwg. #583-145)
1.
Remove any insulation from the installation area, as well as any loose paint, rust and high spots from the pipe.
2.
Remove the backing paper from the inside of the pipe damping jacket (shown in Figure 12 below).
Pipe Diameter
PDJ Part Number
6” (152mm)
583-145-06
8” (203mm)
583-145-08
10” (254mm)
583-145-10
12” (305mm)
583-145-12
*If a pipe damping jacket other than those listed above is
required,
contact the factory for a special order.
TRANSDUCER
WINDOW
4 places
DAMPING MATERIAL
MOUNTING BRACKET