MECH-8
1-8. LISTA Sensitivity Detection
TP
In case of use VFK1409A
F2
: ATF-ERR (VFK1409A),
TP2
: TRG/HSW (VFK1409A),
GND
: GND (VFK1409A)
In case of use VFK1409S
F2
: ATF-ERR (VFK1409S),
TP2
: TRG/HSW (VFK1810),
GND
: GND (VFK1409S)
VTR MODE
PLAY
ADJ. MODE
Refer to below explanation
TAPE
VFM3000LS (DV LISTA)
SPEC.
70 mV /
µ
m to 170 mV /
µ
m
1. Set the camera recorder to VCR mode.
2. Insert the DV Alignment Tape (VFM3000LS) to the camera recorder.
3. Press the button in order of
“ADUB button”
→
“STOP button”
→
“MODE CHK button”
→
“MENU button”
,
to open the
“VCR FUNCTION MENU”
.
4. Select the item
“ADJUST MENU”
and press the SET(STILL) button to open the
“ADJUST MENU”
.
5. Select the item
“ATF GAIN”
and press the SET(STILL) button.
6. Select the item
“ON”
and press the SET(STILL) button, then message
“NOW SERVO ADJUST PUSH MENU
TO RETURN”
appears on screen and press the PLAY button to playback the tape.
7. Select
item
“<1> Sensitivity Measurement”
on the LISTA main menu, and press
“ENTER”
.
8. Then the tape is played back (tape speed : 101.2%) automatically, and message
“1.2% Speed…”
appears on
the screen.
9. Press the ENTER key, and then start measurement of the sensitivity value.
10. Confirm that the sensitivity value is with in specification, after the message
“<<Sensitivity Measurement
Finish>>”
.
Figure 9
<<Sensitivity Measurement Finish>>
Sensitivity 142.95
(mV/um)
Sens.1
138.43 (mV/um)
Sens.1
147.48 (mV/um)
Confirm that the this value in specification
Select at Sensitivity Detection.