Aicure ANUP5255V2
Setting
26
Select a UV irradiation pattern to execute.
1
Enter the “Intensity Selection” screen.
Setting mode
Intensity selection
*1
2
3
4
Special intensity
5
6
Back
Next
Go
▲
(Back)
Moves to the option immediately above the
current option.
▼
(Next)
Moves to the option immediately below the
current option.
SET (Go)
Calls up the “Intensity Selection” screen.
Select “Intensity Selection” with the “
▲
” and “
▼
” buttons, and
press the “SET” button. Then the “Intensity Selection” screen is
called up. (The item is displayed inversely.)
2
With the “
▼
” button and the “SET” button, select from
among the irradiation patterns: UV1 to UV6.
•
Any of the patterns UV1 to UV4 can be selected any
number of times. (It is not allowed when UV5 or UV6
is selected.)
•
When UV5 is selected, no other irradiation pattern
can be selected.
•
When UV6 is selected, no other irradiation pattern
can be selected.
Setting mode
Intensity selection
*1
2
3
4
Special intensity
5
6
Back
Next
Set
▲
(Back)
Terminates “Intensity Selection” and returns to
the “Reference Intensity Setting/Calibration
Setting” screen.
▼
(Next)
Moves the irradiation pattern to 1 to 6.
SET (Set)
Switches the inversely-displayed number
between Selected (with *) and Not Selected
(without *).
Move the highlight bar with the “
▼
” button to display in inverse
video the irradiation pattern number you want to select. Press
the “SET” button and then “*” will be displayed.
To deselect, press the “SET” button once again. “*” will
disappear.
3
After completion of the selection, press the “
▲
” button
(Back) to return to the “Intensity Selection” screen.
Caution:
Summary of Contents for Aicure ANUP5255V2
Page 4: ......