Aicure ANUP5255V2
UV irradiation
44
■
Intensity check
This function adjusts the intensity of UV irradiation to the value selected in “Intensity Selection” in UV
control ON situations. It should be executed in the following cases.
・
Before UV irradiation is performed for the first time.
・
After lamp-change
・
After change of the UV intensity setting
・
After change of the UV intensity pattern
If “Intensity check” is not performed, the intensity error may arise during the initial
irradiation.
1
Select “Intensity check” with the “
▲
” and “
▼
” buttons.
(Figure of the
↓
button section)
Intensity check
*(1)0700
T0003
(2) 1000
T0003
(3) 1500
T0003
(4) 2000
T0003
Back Next Help
▲
(Back)
Returns to the main screen.
▼
(Next)
Goes into the “Intensity Check” screen.
SET
(
Help
)
Press the “
▼
” button to get into the “Intensity Check” screen.
Press the “MANUAL” button to open the shutter.
The patterns (indicated by “*”) selected in “Intensity Selection”
sequentially undergo UV intensity adjustment.
After a given period of time elapses, the current pattern
automatically changes to the next pattern.
2
After the completion of adjustment, switch the “MANUAL”
switch to close the shutter.
* If the UV intensity shown on the LCD has not yet reached the
set value, repeat Steps 2 and 3 until the adjustment is
completed.
3
Press the “
▲
”, “
▼
”, or “SET” button to return to the main
screen.
Caution:
1.3
1.2