Aicure ANUP5255V2
How to replace the lamp
63
8
Install the positive lead wire (long) to the lamp shaft and
secure it with the screw removed in Step 2.
・
Make sure not to mistake the "+" and "-" polarities of the
lead wires when connecting them. Otherwise, the lamp
will not operate.
・
When connecting the "-" lead wire, make sure that the
lamp is not subjected to any load. Otherwise, it may be
damaged.
9
Reinstall the left-hand cover.
11.2
Clearing the total illumination time
(See P. 46.)
Caution:
Negative lead wire
Positive lead wire