Aicure ANUP5255V2
How to replace the Lamp
64
11.3
How to replace the positive lead wire
When replacing the lamp, also replace the positive lead wire if it is damaged.
If operation is continued without replacing a damaged positive lead wire, the unit may break.
Before replacing the wire, turn off (O) the POWER switch, unplug the power cord,
and leave the equipment for at least 30 minutes after turning off the lamp.
1
Before reinstalling the left-hand cover after replacing the
lamp according to “Section 11.1”, remove the six screws
and the right-hand cover.
2
Replace the used positive lead wire with a new one and
correctly wire it.
Make sure to install the new wire in the correct
direction.
3
Reinstall the left and right-hand covers.
When reinstalling the covers, be careful not to let the
lead wire get caught in the covers.
Caution:
Caution:
Caution:
Power supply
Install a new positive lead wire
here.
Right-hand cover