Precautions before use
1) Do not connect the system to a power supply outside the rated voltage/frequency range listed on
the system or in this manual. Using with other than the indicated power could cause burnout.
•
Input Power: Commercial 90 to 264 V AC, 50 Hz/60 Hz
2) Please use the system within the ambient conditions listed below. Using under other than the
indicated conditions could shorten the system lifetime.
•
Ambient temp.: 10 to 40
o
C
•
Relative humidity: No greater than 80%, with no condensation
3) When handling the lamp, do not touch the glass with your bare hands.
If foreign matter adheres to the glass, it could shorten the lamp lifetime or impact curing
performance, or cause the lamp to break. If the lamp is touched with a bare hand or becomes
soiled, wipe clean with alcohol.
4) Make sure the system's rubber legs are set horizontally parallel before use.
Using the system while it is titled, tipped on one side, or upside down could cause the lamp or
main unit to overheat and break. Doing so is also dangerous.
5) The system can be used in two modes: vertical light, and horizontal light. The magnets must be
set in accordance with the installation orientation. Failure to do this could cause the lamp to
shatter, or shorten its lifetime. (See page 10).
6) In order to avoid damage due to overheating, do not use if the area around the system is blocked,
or with two systems back to back. Blocking the bottom or back of the unit could change its internal
cooling conditions, causing damage due to overheating. If two systems are used back to back,
exhaust could cause damage due to overheating.
7) Do not leave the lamp on for prolonged periods while the cover is off. Doing so could cause the
lamp or system to overheat and break.
8) In order to avoid lighting failures due to overcooling, attach the light guide fiber unit to the jack on
the front of the main unit, and make sure it is inserted before use.
9) Do not use the lamp for more than 3,000 hours. Doing so could cause it to break.
Even if the lamp has been used for less time than this, you should always keep a spare lamp on hand.
10) As a rule of thumb, the lamp should be turned on no more than twice per day. Turning the lamp
on more than twice per day could shorten its lifetime.
11) Perform periodic system cleaning about once a month. Dust or other foreign matter could cause
fires or impact lamp cooling.
12) Opening and closing the shutter in rapid succession, vie panel operation or direct operation,
could cause malfunction (minimum shutter open 1 second, minimum shutter closed 1 second)
Warranty and compensation for production
Warranty
Any failures that occur within 1 year (8 hours of use per day) of delivery, under
conditions of normal use, will be repaired free of charge. This does not apply,
however, if the failure was the fault of the user, or due to normal wear or force
majeure, including natural disasters.
Note: If the system is used for 12 or more hours per day, the warranty period is 6 months.
Compensation for Production
We cannot compensate losses due to halted production or defects caused by
system issues.