ANUP5204 Aicure
Irradiation
33
Execute Pattern Irradiation
5
On the operation panel, in the "SHUTTER" section, press
the "AUTO START" button until you hear a beep. The
shutter opens, and irradiation begins in accordance with
the registered pattern, starting with step 0. While the
shutter is open, the "OPEN" LED in the "SHUTTER"
section is lit.
If you would like to check the UV intensity during
pattern irradiation, press the "UV POWER" button.
The value appears in the display.
End Pattern Irradiation
After the entire pattern that was set ends, the shutter closes
automatically, and irradiation ends.
To Forcibly End Irradiation Midway During Pattern Irradiation
Press the "AUTO START" button again until you hear a beep. The
shutter closes, and irradiation ends.
The pattern mode is retained even if the power is
interrupted. It is not necessary to switch to pattern
mode or select pattern numbers each time you turn
on the power.
Canceling Pattern Mode
To cancel pattern mode and switch to normal mode, press and hold the "PATTERN" button until you
hear a beep (at least 3 seconds). The LED above the "PATTERN" button turns off, and the system
switches to normal mode.
Warning
Warning