32
EMC Relevant Information
1.The Intraoral Camera needs special precautions regarding EMC and needs to be installed and put
into service according to the following EMC information.
2.The portable and mobile RF communications equipment can affect the Intraoral Camera.
3.WARNING: The Intraoral Camera should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the Intraoral Camera should be observed to verify normal
operation in the configuration in which it will be used.
4.WARNING: The uses of accessories, transducers and cables, with the exception of transducers and
cables sold by the manufacturer as replacement parts, may result in increased emissions or
decreased immunity of the Intraoral Camera.
Guidance and manufacturer’s declaration - electromagnetic emissions
The Intraoral Camera is intended for use in the electromagnetic environment specified below. The
customer or the user of the Intraoral Camera should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Intraoral Camera uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Intraoral Camera is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies