117
3-5-13 Safety function input (interlock)
T1
T1
T2
T4
T3
T2
T3
ON
OFF
ON
OFF
ON
OFF
ON
OFF
CLOSE
OPEN
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
[Input]
[Output]
READY OUT
MARKING OUT
MARK END OUT
LASER OUT
LASER IN
TRIG. IN
ALARM RESET IN
SHUTTER
OUT *2
Marking interrupted
(Open)
(Close)
ALARM OUT
(Active Low)
SHUTTER B IN
INTERLOCK
SHUTTER A IN
*1 : Control Interlock 1 and 2 on LP-Sxxx-SF/LP-Sxxx-LS1 type with the same flow as the Interlock input above described.
*2 : Shutter output signal has around 200ms to max. 1 sec. delay from each input signal for the shutter open / close
operation time.
The operation behavior of Shutter 2 output on LP-Sxxx-SF/LP-Sxxx-LS1 type is the same as Shutter output.
Item
Time
Remarks
T1
100 ms or more
Keep the ON status for 100 ms or more.
T2
―
Time for data creation or time for laser pumping, whichever is longer.
T3
approx. 20 sec.
The time from turning ON the laser pumping input to completion of pumping.
LP-Sxxx type: approx. 20 seconds / LP-SxxxW type: approx. 15 seconds
T4
2 to 510 ms
Even when the marking is interrupted due to the laser stop input, the mark end outputs.
This signal is a one-shot output. Set the output time on the environment setting screen.
There is a margin for the setting value.
ンㄆㄇㄆㄓㄆㄏㄆ
• If interlock is opened, the laser pumping is turned off and the shutter is closed, regardless of whether or not the laser is
radiating. (Alarm status)
• For release method of alarm, refer to “Release method of alarm” (P.161).
ME-LPS-SSM-8
Summary of Contents for LP-S Series
Page 15: ...15 MEMO ME LPS SSM 8 ...
Page 26: ...26 Chapter 1 Specification ME LPS SSM 8 ...
Page 41: ...41 Chapter 2 Preparation ME LPS SSM 8 ...
Page 78: ...Chapter 3 Connection for External Control ME LPS SSM 8 ...
Page 126: ...Chapter 4 Maintenance ME LPS SSM 8 ...
Page 148: ...Troubleshooting ME LPS SSM 8 ...
Page 169: ...Index ME LPS SSM 8 ...
Page 172: ...172 MEMO No 9000 0066 16V ME LPS SSM 8 ...
Page 173: ...No 9000 0066 16V ...