16
ALARM TEMPERATURE SETTING
This unit is provided with the high temperature alarm. The setting of high temperature alarm is 10
o
C or
15
o
C (The setting value is 2 kinds.) higher than the setting of chamber temperature. The procedure in
table 2 shows the sequence to set the high temperature alarm at 15
o
C higher than the setting of chamber
temperature. Set the figure to 000 at step 5 when setting the high temperature alarm at 10
o
C higher
than the setting of chamber temperature.
Note:
The alarm temperature is set at the factory 10
o
C higher than the setting of chamber temperature.
Table 2. High temperature alarm setting (15
o
C higher than the setting of chamber temperature)
Description of operation
Key operated
Indication after operation
1
-----
The current chamber temperature is
displayed.
2
Press the numerical value shift key
for about 5 seconds.
The
first
digit
blinks.
3
Set the first digit to 1 with the
numerical value shift key.
The
first
digit
blinks.
4 Press enter key.
ENT
The first digit blinks.
5
Set the first digit to 1 with the
numerical value shift key.
When pressed, the figure of settable
digit changes.
6 Press enter key.
ENT
Alarm temperature is memorized and
the current chamber temperature
is displayed.
Note:
The setting is not memorized if the mode change key (PV/SV) is pressed during alarm
temperature setting. The chamber temperature is displayed for 3 seconds when the numerical value
shift is pressed 5 seconds continuously.