Stable temperature control

The MDF-U731M provides precise and 
uniform storage temperatures regardless 
of ambient conditions through 
microprocessor temperature control.

Energy-saving operation

The specially designed compressor 
results in an energy-saving medical 
freezer with superior cooling and quiet 
operation.

Improved security

Standard door lock and independent 
padlock ensure extra security. 
(*padlock not supplied)

Sturdy cabinet construction with 
superior insulation

The integral foamed cabinet structure is 
extra strong and prevents cold air loss.

Manual defrost

Manual defrost freezers provide stable 
temperature control and do not exhibit 
the transient temperature increases that 
can be associated with auto-defrost type 
freezers. The MDF-U731M is therefore 
ideal for storing your important, 
temperature-sensitive samples.

Movable wire shelves allow free 
position adjustment

   
Temperature range
Maximum cooling performance
External dimensions (W x D x H)

*

1

Internal dimensions (W x D x H)
Effective capacity
Net weight (approx.)
Exterior
Interior
Insulation
Temperature control
Compressor
Defrosting system
Refrigerant
Shelves (max. load)
Access port
Casters

Alarm and safety

Accessories

Control panel

Options

Performance Data

Pull-down characteristics

Pull-up characteristics

 

-20˚C to -30˚C (1˚C increments)

-30˚C (Ambient temperature: 35˚C)

770 x 830 x 1955 (mm) / 30.3 x 32.7 x 77.0 (inch)
650 x 700 x 1520 (mm) / 25.6 x 27.6 x 59.8 (inch)

690 liters (24.4 cu.ft.)

152 kg (335 lbs.)

Painted steel
Painted steel

Rigid polyurethane foamed in-place

Electronics controller

 Hermetic type (400 W)

Manual defrost

HFC

4 adjustable wire shelves (50 kg/110 lbs.)

ø 30mm, 2 locations (left and top)

4 casters with 2 adjustable feet

High / Low temperature alarm, Power failure alarm, Door alarm, 

Compressor temperature alarm, Self-diagnosis function, 

Remote alarm contact (DC30V, 2A), Memory backup

1 scraper, 1 set of key, 2 large & 2 small nylon clips (for temperature recorder)

Chamber 1/2H Central Air temperature (˚C)

Conditions: 

AT 35˚C, no load, measured at center of chamber

Time (hr.)

Chamber 1/2H Central Air temperature (˚C)

7-day circular 

recorder

MTR-G85C-PE
MTR-G85A-PA

32-day strip 

chart recorder

MTR-4015LH-PE
MTR-4015LH-PA

Recorder mounting bracket

MPR-S7-PW

Recorder mounting bracket

MPR-S30-PW

Wire shelf (3 sets)

MDF-T07ST-PW

Storage case (2 sets)

MDF-T07SC-PW

*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30˚C with no load). In order to use the freezer 

at a stable temperature for a long time, it is recommended that the temperature be set to at least 5˚C higher than the indicated lowest temperature. In addition, 
depending on the usage conditions, it may not be possible to reach the indicated lowest temperature.

Caution: 

Panasonic guarantees the product under certain warranty conditions. Panasonic in no way shall be responsible for any loss of content or damage to content. 

•Appearance and specifications are subject to change without notice.

*1 Exterior dimensions of main cabinet only - see dimension drawings showing handles and other external projections.

We design and manufacture the product in accordance with ISO9001 requirements.

Time (hr.)

–40

–20

0

20

–30

–10

10

30

0

1

6

2

9

8

10

7

4

3

11

5

–40

0

–30

–20

30
20

–10

10

0

5

2

6

3

4

8

1

7

Printed in Japan   188-2013-06

http://panasonic.net/healthcare/biomedical/

650 (25.6)

770 (30.3)

1491 (58.7)

1520 (59.8)

830 (32.7)

1955 (77.0)

89 (3.5)

700 (27.6)

362.5 (14.3)

362.5 (14.3)

362.5 (14.3)

10 (0.4)

MDF-U731M-SE
MDF-U731M-SA

230/240V, 50Hz
115V, 60Hz

Model No.

Reviews: