List of Error Codes and Description 

 
 

A3 Changer not moved to next shelf error 
A5 Changer not fully loaded or returned error 
A6 Autochanger solenoid failure 

A –  
Automatic Magazine 
Changer 

A7 Unit door is open 
B0 Badge not pushed up error 
B1 Badge not pushed down error 

B –  
Badge Inserter 

B2 Badge not pushed down error 

M- 
Magazine Driver 

M0 Magazine not moved to next slot. 

N0 NCR lamp error 

N- 
Number Code 
Reader 

N2 NCR Parity error 

R- 
RS232C Interface 

R2 RS232 TX data ACK time out 

 

 

 
A-Type Error General Troubleshooting 
Each of the specific A-type errors discussed below causes the UD-710 automatic 

IRRADIATOR

 to cease 

automatic operation.  No dosimeters are read after an A-type error occurs until the problem is resolved.  This type 
of error occurs for one of two reasons.  First, a timing sensor did not detect a signal within the specified time limit 
even though the upper sensor detected the top shelf of the Automatic Magazine Changer.  Second, the shelf 
location in the 

IRRADIATOR

 memory was the first shelf, even though the upper sensor did not detect the top shelf.  

Troubling this type of error consists of removing any obstacle, checking the mechanism or replacing faulty parts 
according to the Service Manual. 
 
A3 Errors 
This is a timing sensor error which occurs when a signal is not received within a specified time limit of moving one 
shelf up or down even though the rack driving motor is operating.  This error is usually caused by something 
blocking the magazine rack, failure of the driving motor, failure of the sensor, or failure of the printed circuit (PC) 
board of the driving circuit.  This error may be resolved by removing any blocking material, checking the 
mechanism or its control circuits or replacing any faulty components. 
 
A5 Errors 
An A5 error is produced by one of two timing sensor problems.  First, an A5 error may be produced when the 
magazine end sensor of the loader does not detect a signal within the specified time limit even though the 
magazine driving motor is operating in reverse and the magazine sensor of the reader detected the magazine 
head.  An A5 error may also be produced when the magazine sensor of the 

IRRADIATOR

 does not detect a signal 

within the specified time limit even though the magazine driving motor is operating.  The usual causes of this type 
of error are something blocking the path of the magazine, failure of either the sensor of the loader or reader, or 
failure of the driving motor.  This error may be corrected by removing any obstacle, and adjusting or replacing the 
faulty components. 
 

Summary of Contents for UD-794D

Page 1: ...anual Operation Manual Operation Manual TLD IRRADIATOR TLD IRRADIATOR TLD IRRADIATOR TLD IRRADIATOR MODEL MODEL MODEL MODEL UD UD UD UD 794D 794D 794D 794D 2011 12 OperationManualVersion10 05_EU UD 794 Panaasonic ADVANCEDTLDSYSTEM ...

Page 2: ...dforirradiationofPanasonicTLDsonly IndoorUseOnly Donotexposetorain heavymoisture drippingandsplashing Indoo AlwaysTransportUpright Pleasenotethefollowingtwolabelsinsidethemechanism ENTANGLEMENTHAZARD ENTANGLEMENTOFHANDMAYRESULTINSEVEREINJURY DONOTTOUCH DISASSENBLYPROHIBITED DONOTOPENTHISCOVER TRAINEDSERVICEPERSONELLONLY SPECIFICATIONSWITHINTHISMANUALARESUBJECTTOCHANGEWITHOUTNOTICE SAFETY INSTRUCTI...

Page 3: ...ation 6 Operation 10 MainScreen 13 FileMenu 14 New BatchMenu 14 ViewMenu 15 BadgeResultsMenu 15 StatusLogMenu 16 ReportsMenu 17 ModeMenu 18 LoaderMenu 18 ChangerMenu 22 ParametersMenu SystemParametersMenu 27 IndividualDoseSettingMenu 28 RS232OptionsMenu 29 ErrorCodes 30 ...

Page 4: ...te10magazinesof50TLDbadges Descriptionofparts Frontview Mainkeyswitch Insert themainkeyandturnclockwise ACpowerindicatorlight Whenpoweronthenthislight illuminateRED Leftdoorpanel Asourceenclosureexistsintheinterior Donotopenthisdoor Leftdoor handlewithkeyhole Right door ThisdoorisopenedthenTLDBADGEtrayssetting Donotopenthedoorthenoperationgstate IfopentheRightdoor theninnerACpowerline isshutdown R...

Page 5: ...PowerSW Equipment UD 794 powerSW ACInput AC220V 240V50 60Hz FuseT6 3A ACOutput 3port AC110V 120V50 60Hz Fuse3 15A InsidemotorFuse 1A ParallelI FcableforDIOboardinsertedinthePCextendedPCIport GNDcableforconnectbetween PCandUD 794equipment ...

Page 6: ...npowerSW ACinletport AC220V 240V50 60Hz mainfuseBox T6 3A 2pieces UD 794powerSW Fuse forinternalACmotor 1A FuseforACoutlet 3 15A ACoutletport 3Port forPC Display monitor CAUTION AC110V 120Vtotal3 15A ModelNo IRRADIATOR2NCE TLDIRRADIATOR ...

Page 7: ...Monitor Nootherequipmentshouldbeattached tothese Operating Temperature 5 35 20 70 RH Powersource AC220V 240V 50 60Hz Dimensions 1000mmX550mmX1120mm WxDxH Weight Approx 220kgwhichincludessourceandshield Sound level 75dB 1m Attachment ACcableforAC220V 240V 1piece ACcableforUD 794ACoutlet 2pieces DIOboardforextendedPCIportof PC 1piece DIOI Fcable 1piece GNDcable 1piece Operationgmanual 1piece Doorkey...

Page 8: ...iththeIrradiatorsystem Thisdevicecontainsasourceofionizingradiation OperationislimitedtoQualifiedPersonel SetequipmentonthehorizontalLevel hard floor Equipmentsidespace seebelowfigure Equpmentsidespacearea A LeftsideandRightsidespaceisminimum20mm B Rearpanelspaceisminimum100mm C Toppanelspaceisminimum0mm A A B C ...

Page 9: ...ythattheboardsinthePCexpansionbusareproperlyinsertedand seatedintheirslots 5 Connect theNationalInstrumentsDIOI Fcableandgnd cablefrom theUD 794casetotheslotinthePC expansionboard See belowfigure 6 Connectthetwo50 pinribboncableassembliestothe50 pinheadersH1andH2onthedigitalI Oboard InsurecorrectPin1orientationwithPin1oftheribboncableassemblies Brownispin1 ...

Page 10: ...eUD 794 SelectONfromtherearACinput turnontheinternalpowereswitchandfinallyturnthe mainkeyswitchlocatedonthetop rightsideoftheirradiatorclockwise 9 PowerontheDisplaymonitor andPC NowstartUD 794programbyclickingontheIrradiatorprogramicon TheACpowerindicatorlightwillilluminateatthistime Followthesimplemenuinstructionstooperatethe UD 794 FromUD 794 FromUD 794 PC Expansionboard GNDcable DIOI Fcable ...

Page 11: ...entryischeckedforvalidrangesandlimitsuponentry Invalidinputdatawillbeignoredandthecomputerscreenwilldirectyoutoenterdatainavalidrange TheprogramstartswiththePanasonicTitlescreen asshownbelow Softwareversionsmaybedifferent Theoperatorispromptedtoentertheproperpasswordandoperatorname Afterashortdelay duringwhichthemotorsareinitialized themainsectionoftheprogramwillthenbeginatthe Mainscreen ...

Page 12: ...exposuretimeand oranexposuredosearethenenteredforeachtrayorsingleinputdependingon DosagesandTimeMode togglebutton Ifthetogglebuttonissetto SingleInput theoperatorcansettimeor doseforall10trays Ifthetogglebuttonissetto MultipleInput theoperatorshouldenterindividualtimeanddose foreachtray Theoperatorusesthemouseorkeyboardtochangeexposure Time orexposure Dose If exposuretimeischanged exposuredoseisch...

Page 13: ...eirradiationprocessiscompleted andtheymayremovethe magazine s fromtheIrradiator 5 Theabovesequenceofstepsisthenrepeateduntilalldesiredbadgesareirradiated 6 Theautomaticirradiationprocessmaybestoppedwithsomeerrors Theoperatormayusethemouseto selectthe Stop buttontostoptheprocess ResetMotors buttonmaybeusedtoresetallmotorstohomepositionfollowingadjustment Alldrop downmenusareexplainedinthefollowingp...

Page 14: ...Logmenuitemisfordisplayingeventlogs View BadgeResultsMENU View BadgeResultsscreenshowsbadgeprocessingresults Itshowstime traynumber slotnumber badge number type error doseexposure doseexposurerate anddoseexposuretime Itcanbeprintedorsavedas anotherfilefromthemenu View StatusLogMENU View StatusLogscreenshowsaseriesofevents Itshowsdate time andeventdescription Itcanbeprintedor savedasanotherfilefrom...

Page 15: ...rCheck LoaderTestMenuisthemenufromwhichIrradiatorirradiatormechanicsfunctionsare testedandotherdiagnosticfunctionscanbeinvoked TheMode ReaderCheck LoaderTestMenuscreenandits operationandfunctionareexplainedbelow PleasedonotoperatewithoutguidancefromaPanasonicengineer TheTestMenus canbeopenedbeforeandduringoprtationoftheIrradiator allowing theusertoobservethe stateoftheIndicators forreportingtoPana...

Page 16: ...derfunctions NumberCodeReaderfunctionsareusedtotestthenumbercodereader mechanics TheBadgeNumberCodeReadMenuisexplainedindetailintheBadgeNumberCodeReadMenu description Mode ReaderCheck LoaderTestMENU MagazineMotorFunctions ThesefunctionsallowtheoperatortotestandexercisetheIrradiatorIrradiatorsystemmagazinemotor horizontal drive mechanics AlignMagMotorbutton Thisfunctionalignsmagazinemotor MagStartb...

Page 17: ...onitorsthestatusofthemagazineendsensorRSR7 Whenthemagazinehasindexed tothelastslot slotnumber50 thissignalwillchangefromlogic0tologic1toindicatetheendofthemagazine tray MagAlignedindicator MonitorsthestatusofthemagazineslotsensorRSR10 Whenamagazineisindexed toaslotnumber thissignalwillchangefromalogic0toalogic1ifthatslot canbeanyoneoftheslots 1 50 is properlycenteredabovethebadgepush upmechanism M...

Page 18: ...RLampOff Oncontrol Thisoutputsignalturnsonthelamparrayinthenumbercodereader On position turnsonthelamparraysothatthepunchednumbercodedatacanberead Off positionturnsoffthelamparray ReadNCRbutton Thisfunctioncausesabadgetotraveltothenumbercodereadpositionwherethebadge numbercodeisread Thebadgenumbercodedataiscontinuouslydecoded formattedanddisplayedinthe NCRwindow Then thebadgeismovedbacktothehomepo...

Page 19: ...utilizesthemousetoselect Mode Reader Check Changer menu Thechangertestscreenthenappears Solenoidcontrol Atogglefunctionwhichengages out ordisengages in themagazinedrive horizontaldrive solenoidwhichinturn engagesthemagazinedrivegear Direction Atogglefunctionwhichdeterminesthedirectionofthehorizontalandverticaldrivemotors Down Outmeanstheshelf feedingunitwillgodownandthemagazinedrivemotorwillrevers...

Page 20: ...orsifthemagazineisathomeposition logic1 ornot logic0 TrayLoadedindicator Aninputsignalwhichmonitorsifamagazineispresentornotonashelfinthe Logic1 meansamagazineispresent whilelogic0meansnomagazineispresentonagivenshelf AtBottomindicator Aninputsignalwhichshowsifthechangerisonbottom logic1 ornot logic0 AtTopindicator Aninputsignalwhichshowsifthechangerisontop logic1 ornot logic0 TrayAlignedindicator...

Page 21: ...theCS 137sourceforthehighandlowattenuationsettings Thecalibrationdateallowsforreal timecompensationfortheradioactivedecayoftheCs 137source Theseparametersareinitiallysetby totheexactdoseratemeasuredatthebadgeirradiationpositionbytraceable probesandcalibratedmeasurementequipment SetUnattenuated High Travel Thisfunctionallowsthemanagertoenteranunattenuatedtraveldoseforthe source inunitsofmSv SetAtte...

Page 22: ...IGHtravelingdoseis 0 281mSv sx0 0732s 20 6µSv SourceCalibrationDate MM DD YYYY Thisfunctionallowsthemanagertoenteranewsourcecalibration dateinthe militaryformatMM DD YYYY MMis1to12forthemonth DDis01to31forthedate YYYYistheyear Itisrecommendedthat theirradiatoriscalibratedrelativetoaPromaryorSecondaryStandardSourceunder knownconditionssuchas 1 IrradiationFrontBuild up 2 PhantomorFree in Air 3 Insid...

Page 23: ...emouseto highlight IrradiatorIDNumber andthenewnumberisthenentered ParityChecking Thisisatogglefunctionwhichalternatelyenables Enabled ordisables Disabled automaticparitycheckingofthebadgenumbercode PulseDelayMotor HuntTime Thisfunctionallowsthepulsedelayofthesteppermotorstobeset Thiscritical parameterisnormallysetbyfactorypersonneltoappropriatevalues Sincepropersystemoperationanderror checkingisb...

Page 24: ...usedtoselecttray doseexposuresettingorindividualdoseexposure settingforeachbadgeintheselectedtray LoadfromFile buttonisusedtoretrieveasavedirradiationtraypattern SavetoFileAs buttonisusedtosaveanirradiationtraypattern Save buttonisusedtosavetocurrentparametr inifile Exit buttonisusedtoexittotheMainscreen CAUTION Thismenuover ridesthedosesettingsontheMainScreen Page10 ...

Page 25: ...thisoptionisselected theIRRADIATORisplacedinthe Basic mode DataaretransmittedthroughtheRS 232Coutputportto theexternaldeviceonlyifa receivedtextdata signal RXD isreturnedbythedevicethatacknowledges ACK thedatawere received AnACKsignalisreturnedforanevenparity Otherwise anot acknowledge NAK signalisreturned IOQueueWaitTimeout DeterminesIOqueuewaittimeoutinsecondswhentheIRRADIATORtransmitsdatathroug...

Page 26: ...hemechanismorreplacingfaultyparts accordingtotheServiceManual A3Errors Thisisatimingsensorerrorwhichoccurswhenasignalisnotreceivedwithinaspecifiedtimelimitofmovingone shelfupordowneventhoughtherackdrivingmotorisoperating Thiserrorisusuallycausedbysomething blockingthemagazinerack failureofthedrivingmotor failureofthesensor orfailureoftheprintedcircuit PC boardofthedrivingcircuit Thiserrormaybereso...

Page 27: ...persensordoesnotdetectasignal withinthespecifiedtimelimit eventhoughthespecifiednumberofpulsesweretransferredtothepulsemotorfor badgeinsertionintothenumbercodereaderhousing Theprobablecausesofthiserroraresomethingblocking thepathofthebadge themagazinepositioningisincorrect failureofthepulsemotororsensor orfailureofthe mechanismcontroller Thiserrormaybecorrectedbyfollowingthetroubleshootingsuggesti...

Page 28: ...anN2erroroccurs IfanN2erroroccurs aquestionmark willbeprintedincolumn8ofthe printoutandthebadgewillnotberead AutomaticoperationwillcontinueuntilthreesuccessiveN2errorsoccur atwhichtimeautomaticoperationisinterrupted Theprobablecausesofthiserrorincludeamispunchofthe parityholeinagivenrow adecreaseinthesignalleveloftheopticalarrayduetoabuildupofdust failureofthe readingcircuits orfailureofthesensors...

Page 29: ...26 PleaseaddressanyenquirieswithintheEuropeanAreato PanasonicIndustrialEuropeGmbH UK WilloughbyRoad Bracknell Berks RG128FP UNITEDKINGDOM Tel 01252876292 Tel 07836617737 ...

Reviews: