Chapter 3 Troubleshooting
003-322 IISS-ESS Communication Failure 3
BSD on : CH12.2A
[Detection Conditions]
SCOPE-FD sending error that will be detected by the
Controller (Incorrect Packet No.)
ACK cannot be received after resending it twice. (The
Packet No. of the message Packet that has been sent is
not correct.)
[Actions]
Turn the power On/Off and perform the following if
the problem persists.
(1) Check the connection of the IIT/IPS (PL16.3)
Connection Harness.
(2) Software matching error. Perform re-installation of
the software.
(3) If the problem persists after the above actions,
replace the IIT/IPS (PL16.3).
003-323 IISS-ESS Communication Failure 4
BSD on : CH12.2A
[Detection Conditions]
SCOPE-FD sending error that will be detected by the
Controller (Incorrect message length)
ACK cannot be received after resending it twice. (The
message length of the message packet that has been
sent is not correct.)
[Actions]
Turn the power On/Off and perform the following if
the problem persists.
(1) Check the connection of the IIT/IPS (PL16.3)
Connection Harness.
(2) Software matching error. Perform re-installation of
the software.
(3) If the problem persists after the above actions,
replace the IIT/IPS (PL16.3).
003-324 IISS-ESS Communication Failure 5
BSD on : CH12.2A
[Detection Conditions]
SCOPE-FD sending error that will be detected by the
Controller (Check Code Error)
ACK cannot be received after resending it twice. (The
Check Code of the message packet that has been sent
is not correct.)
[Actions]
Turn the power On/Off and perform the following if
the problem persists.
(1) Check the connection of the IIT/IPS (PL16.3)
Connection Harness.
(2) Software matching error. Perform re-installation of
the software.
(3) If the problem persists after the above actions,
replace the IIT/IPS (PL16.3).
003-325 IISS-ESS Communication Failure 6
BSD on : CH12.2A
[Detection Conditions]
SCOPE-FD sending error that will be detected by the
Controller (Parity Error)
ACK cannot be received after resending it twice. (A
parity error is found in the hardware at the IISS side.)
[Actions]
Turn the power On/Off and perform the following if
the problem persists.
(1) Check the connection of the IIT/IPS (PL16.3)
Con
nection Harness.
(2) Software matching error. Perform re-installation of
the software.
(3) If the problem persists after the above actions,
replace the IIT/IPS (PL16.3).
003-326 IISS-ESS Communication Failure 7
BSD on : CH12.2A
[Detection Conditions]
SCOPE-FD sending error that will be detected by the
Controller (Flaming Error)
ACK cannot be received after resending it twice. (A
flaming error is found in the hardware at the IISS
side.)
[Actions]
Turn the power On/Off and perform the following if
the problem persists.
(1) Check the connection of the IIT/IPS (PL16.3)
Connection Harness.
(2) Software matching error. Perform re-installation of
the software.
(3) If the problem persists after the above actions,
replace the IIT/IPS (PL16.3).
003-327 IISS-ESS Communication Failure 8
BSD on : CH12.2A
[Detection Conditions]
SCOPE-FD sending error that will be detected by the
Controller (Overrun Error)
ACK cannot be received after resending it twice. (An
overrun error is found in the hardware at the IISS
side.)
[Actions]
Turn the power On/Off and perform the following if
the problem persists.
(1) Check the connection of the IIT/IPS (PL16.3)
Connection Harness.
(2) Software matching error. Perform re-installation of
the software.
(3) If the problem persists after the above actions,
replace the IIT/IPS (PL16.3).
Summary of Contents for Workio DP-C401
Page 4: ...This page is intentionally left blank ...
Page 5: ...Chapter 1 General ...
Page 22: ...1 16 Chapter 1 General 1 3 Type of Paper ...
Page 23: ...1 17 Chapter 1 General ...
Page 24: ...Chapter 2 Maintenance ...
Page 63: ...Chapter 3 Troubleshooting ...
Page 118: ...3 44 Chapter 3 Troubleshooting I F PWB PL9 1 MCU PWB PL13 1 ...
Page 445: ...Chapter 4 Parts List ...
Page 548: ......