PGC, Inc.
Appendix B; SmartPad version 0.503
Steady State & Programmable
November 2005
B-25
SEGMENT 07:
We now wish to ramp back down to our original conditions of 25 deg. C and 45% RH in 40
minutes.
SEG TIME
AIR SET POINT
RH SET POINT
01
00 00 01
25.0
45.0
02
00 30 00
25.0
45.0
03
00 45 00
25.0
55.0
04
00 30 00
25.0
55.0
05
01 20 00
35.0
60.5
06
00 15 00
35.0
60.5
07
00 40 00
25.0
45.0
If it is desired to repeat this cycle, loop from segment 07 to segment 02. If it is desired to hold
the conditions, loop from segment 07 to segment 07. It is advisable to note the start, loop from
and loop to segments on the programming sheet; for this example program: start at segment 01,
loop from segment 07 and loop to segment 02 or 07 as selected. This information will be entered
on the "PROGRAM INITIALIZATION" screen.
This should give the operator the necessary information to write a simple program. After water
has been turned on, wait until reservoir is filled before turning on the conditioner. The pump and
water heater should NEVER be powered without water in the system.
ACTUAL LOADING OF THE PROGRAM
Set the Segment Time, Dry-Bulb Temperature and RH as follows:
NOTE
- If lock out option "ALL" has been chosen from the configuration options screen, there
will be no "TAB" function (F1) and no flashing cursor since you are allowed to view but not
change the program.
1.
From the "METER" screen press the "PROG" key, the display should change to the
"PROGRAM" screen.
2.
Segment 00 should show on the display, depress "NEXT" or 01 on the number pad. The
display will now show the data for segment 01.
3.
Notice that the data entry cursor is presently located in the Segment field. Data can only
be entered into a field if the cursor is located in that field. Press the "TAB" key once, the first
position of the "time" should start flashing.
4.
On the number pad depress 00 00 01. The flashing portion of the display should have
moved all the way through the time indication and back to the beginning. It is necessary to enter
a digit for each character in every field that is to be changed. The data will be entered into
memory as soon as the last digit is entered. If the time is correctly entered, depress the "TAB"
key; if not simply reenter the correct time since the first digit is still flashing.
5.
After pressing "TAB", the first digit of the "air set point" should be flashing. On the
number pad depress 250. It is not necessary to enter the decimal point. The flashing portion of
Summary of Contents for 9354-4250
Page 2: ...PGC Inc Horizontal 500 1000 CFM June 2006 Parameter Generation and Control Inc 2 ...
Page 42: ...PGC Inc Appendix B SmartPad version 0 503 Steady State Programmable November 2005 B 1 ...
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......