I179EN - 1021
7/16
PARATRONIC - Rue des Genêts - Zone Industrielle - 01600 REYRIEUX - FRANCE
Tel.: +33 (0)4 74 00 12 70 – Fax: +33 (0)4 74 00 02 42 - E-Mail: info@paratronic.fr - Web:
http://www.paratronic.com
5.3 SENSOR:
116
35
Ø21.5
Ø7
Ø10
LH(C)
SFP
6
INSTALLATION
The LH(C/M) sensor can, depending on the options selected, be supplied with (LHC) or without
(LHM) an antenna, and with or without a mounting bracket (ref. “LH EQUERRE”). For the LHC, the
antenna supplied as standard is a “whip” antenna; as an option, a “pad” antenna can be supplied
(ref. “LH ANTGAL”). Therefore, different mounting and assembly possibilities are available
according to the equipment and options selected.
6.1 PRECAUTIONS FOR INSTALLATION AND USE:
The recorder module (LH(C/M)) and its associated probe (SFP) are measuring instruments.
Therefore, they must be handled with caution and care. All transportation must be carried out in the
original packaging.
Products must not receive shocks and must not be subjected to severe pressure or temperature
variations.
The probe measurement cell is an active and sensitive component. For this reason, the probe must
be set up (submerged) with caution. Pressure must never be exerted on the cell with a tool, and the
cell must never be cleaned with a corrosive substance. In the event of fouling, it is recommended to
simply rinse with water.
The LH(C/M) recorder module makes an atmospheric pressure measurement. The pressure
measuring port is located on the bottom of the module, next to the connector intended for the
sensor. Please ensure that this port is never blocked.
Always respect the temperature and pressure operating limits associated with the model used
(refer to the “SPECIFICATIONS” section).
LH (C)
SFP