5
TROUBLESHOOTING
In the event of faults that are not listed in this chapter, or if the
suggested procedure does not correct the fault, contact the
PARI GmbH Service Center.
Fault
Possible cause
Procedure
The
rechargeable
battery is
plugged into
the
compressor,
but the
compressor is
not working.
The battery is flat.
Connect the power
adapter to the
rechargeable battery to
charge it.
There is a fault in the
compressor or the
rechargeable battery.
Reset the rechargeable
battery:
•
Switch the com
pressor off.
•
Disconnect the bat
tery from the com
pressor [see also:
3.3 Disconnecting
the rechargeable bat
tery from the com
pressor].
•
Wait at least
10 minutes before
operating the re
chargeable battery
again.
•
If the problem per
sists, try charging the
battery [see also:
3.1 Charging the re
chargeable battery].
•
If the problem still
cannot be corrected,
contact Technical
Service [see also:
7.4 Contact].
All 4 LEDs are
flashing
– The battery is over
loaded (> 3.5 A).
– The battery is over
heated.
– The battery is flat
(deeply discharged).
en
– 25 –