NEBULISED ANTIBIOTICS

 USER GUIDE

TABLE OF CONTENTS

Patient Contact

 6

Aerosol Delivery of Inhaled Antibiotics 

 6

Supplies Checklist

  

7

Handwashing Technique 

 8

Aerosol Therapy Set-Up –  
Nebuliser with exhalation filter

 10

Aerosol Treatment

  

12

Cleaning the Nebuliser 

 14

Disinfecting the Nebuliser  

15

Easycare cleaning aid 

 16

Frequently Asked Questions  

18

Summary of Contents for eFlow rapid

Page 1: ...A step by step guide for patients using the eFlow rapid eFlow rapid USER GUIDE FOR NEBULISED ANTIBIOTICS www pari com...

Page 2: ...I reusable nebulisers available for aerosol delivery of inhaled antibiotics but this booklet deals solely with the eFlow rapid This booklet acts as a supplement to not replacement for the Instructions...

Page 3: ...sol Delivery of Inhaled Antibiotics 6 Supplies Checklist 7 Handwashing Technique 8 Aerosol Therapy Set Up Nebuliser with exhalation filter 10 Aerosol Treatment 12 Cleaning the Nebuliser 14 Disinfectin...

Page 4: ...41G0500 Replacement Filter Pads Part No 041B0522 pack of 30 Part No 041B0523 pack of 100 PARI Filter Valve Set Replacement Filter Pads This is the eFlow rapid nebuliser system AEROSOL DELIVERY OF INHA...

Page 5: ...to handling supplies and medications Wet hands thoroughly with warm water Rinse hands thoroughly under running water Use antibacterial soap if possible Dry with a clean towel Thoroughly wash hands for...

Page 6: ...the nebuliser Replace the medication cap Read and understand the PARI nebuliser instructions for use 5 8 9 6 10 11 7 You should have 1 x Mouthpiece without valve or BABY mask for young children babies...

Page 7: ...enever possible ALTERNATIVELY if using the PARI BABY mask place the mask comfortably over the mouth and nose of the patient 1 4 2 5 6 3 Monitor for side effects If you experience any of the following...

Page 8: ...r Wash all parts of the nebuliser in this solution Rinse all parts thoroughly after washing and place between two clean towels or cloth to dry whilst keeping dust free Do not use kitchen towel All par...

Page 9: ...wear During its entire period of use an aerosol head should not be back washed more than 25 times with the easycare cleaning aid Backwash procedure Start the backwash by pressing the start button on t...

Page 10: ...an optional accessory that will minimise aerosol escaping into the room This may be important if you don t want family members or anyone other than the patient inhaling the antibiotic aerosol mist Ex...

Page 11: ...20 www pari com PARI Medical Ltd The Old Sorting Office Rosemount Avenue West Byfleet Surrey KT14 6LB E Mail infouk pari eu Phone 44 0 1932 341122 Fax 44 0 1932 341134...