Fault
Possible cause/Remedy
The controller cannot be switched on
(no beep, no orange or green light
signal).
- To switch on, press the ON/OFF button
for approx. 2 seconds.
- Is power plug plugged into mains and
the device (for mains operation)?
- Are batteries inserted correctly?
- Check the charge level of the batteries.
After switching the controller on, no
aerosol is generated, and/or the
controller switches off again after
a few seconds.
- Has the medication reservoir been filled
with medication?
- Check the connection between the
controller and the nebuliser handset.
The controller switches off even
though there is still medication left in
the medication reservoir.
If there is still more than 1 ml in the
medication reservoir of the eFlow
®
rapid
nebuliser handset:
- Check the charge level of the batteries.
- Was the nebuliser handset held
horizontally throughout the inhalation
session? Hold the nebuliser handset
horizontally.
- The controller switches off
automatically after 20 minutes.
Carry out inhalation to the end by press-
ing the ON/OFF button again, so that
you can inhale the prescribed dose of
medication.
If the maximum operating time for each
inhalation session is exceeded
repeatedly, the aerosol head must be
replaced.
Info:
Note that about 1 ml of medica-
tion cannot be nebulised and
remains unused in the medica-
tion reservoir of the
eFlow
®
rapid nebuliser handset;
this residue must be discarded.
This is intentional and does not
represent a malfunction.
With medication-specific nebu-
liser handsets (
Tolero
®
,
Zirela
®
and
Altera
®
) there is no signifi-
cant volume of residue left in
the medication reservoir.
The controller does not switch off
automatically even when no more
aerosol is being produced and there
is no more than about 1 ml of residue
left in the medication reservoir of the
eFlow
®
rapid
nebuliser handset.
The treatment session has been com-
pleted successfully and the controller can
be switched off by pressing the ON/OFF
button.
Summary of Contents for eFlow rapid
Page 6: ...2 eFlow rapid 2021 04...
Page 36: ...32 eFlow rapid 2021 04...
Page 66: ...62 eFlow rapid 2021 04...
Page 69: ...eFlow rapid 2021 04 65 el 1...
Page 70: ...66 eFlow rapid 2021 04...
Page 75: ...eFlow rapid 2021 04 71 el 9a Controller Controller 5a 9a 5a...
Page 76: ...72 eFlow rapid 2021 04 3 6 7 60 3 6 7...
Page 77: ...eFlow rapid 2021 04 73 el 8 8 Controller 8 8 6 ml...
Page 78: ...74 eFlow rapid 2021 04...
Page 80: ...76 eFlow rapid 2021 04 Controller Controller 83 eFlow rapid 1 ml 1 ml...
Page 81: ...eFlow rapid 2021 04 77 el 78...
Page 82: ...78 eFlow rapid 2021 04 5 3 5 37 C 5 5 3 easycare 2 easycare...
Page 83: ...eFlow rapid 2021 04 79 el 6...
Page 84: ...80 eFlow rapid 2021 04 5 81...
Page 87: ...eFlow rapid 2021 04 83 el 6 Controller ON OFF 20 ON OFF...
Page 89: ...eFlow rapid 2021 04 85 el easycare easycare easycare easycare...
Page 90: ...86 eFlow rapid 2021 04...
Page 94: ...90 eFlow rapid 2021 04 EN 60601 1 eFlow rapid II BF IEC 60529 IP IP 21...
Page 96: ...92 eFlow rapid 2021 04 10 BF CE 93 42 13 2005...
Page 97: ...eFlow rapid 2021 04 93 el 11 PARI PARI PARI PARI...
Page 140: ...136 eFlow rapid 2021 04 Gakktu r skugga um a allar tengingar s u ttar og a lyfh lfi s l st...
Page 190: ...186 eFlow rapid 2021 04...
Page 250: ...246 eFlow rapid 2021 04...
Page 281: ......