eFlow
®
rapid
21
en
Even when the instrument dishwasher is
equipped with a drying function, ensure
that no residual moisture remains in the
nebuliser handset. Remove residual mois-
ture by shaking the parts and allow them
to air dry completely.
Possible alternative: Chemical
disinfection
In general, disinfectants based on quater-
nary ammonium compounds are suitable
for disinfecting the nebuliser handset,
including the aerosol head.
When choosing the chemical, please
ensure that it is suitable for disinfecting
medical devices made from the specified
materials (see "Materials used", page 28).
Follow the instructions for use of the disin-
fectant, particularly the dosage informa-
tion and instructions for safe use.
The effectiveness of this method has been
tested with a 2% solution of Bomix plus
(Bode) for an application time of
5 minutes.
Controller, connection cord
and power adapter
Cleaning
Clean the controller, the connection cord
and the power adapter as necessary with
a clean, lint-free, damp cloth.
Allow the controller, the connection cord
and the power adapter to dry completely.
Disinfection
Disinfect the controller, the connection
cord and the power adapter immediately
after cleaning. Use an alcohol-based,
standard commercial disinfectant wipe for
disinfecting. To ensure safe application,
follow the instructions for use of the disin-
fectant wipe. Effectiveness has been
tested using Bacillol Tissues and
Clinell Wipes.
Allow the controller, the connection cord
and the power adapter to dry completely.
Storage
Between applications, particularly for pro-
longed breaks in treatment, keep the
nebuliser handset in a dry, dust-free place
(e.g., nebuliser handset bag).
Pack the controller, the connection cord
and the power adapter in the case
provided for this purpose.
NOTE:
Do not hold controller, the connection cord
or the power adapter under running water
and do not use liquid cleaning agents. If
liquid gets into these components, the
electronics may be damaged and the
device may no longer function properly. If
liquid does get into them, contact your
local service partner immediately.
NOTE:
Leaking batteries may cause damage to
the controller. Therefore, remove the bat-
teries if the nebuliser system is not
expected to be used for a relatively long
period of time.
Summary of Contents for eFlow rapid
Page 6: ...2 eFlow rapid 2021 04...
Page 36: ...32 eFlow rapid 2021 04...
Page 66: ...62 eFlow rapid 2021 04...
Page 69: ...eFlow rapid 2021 04 65 el 1...
Page 70: ...66 eFlow rapid 2021 04...
Page 75: ...eFlow rapid 2021 04 71 el 9a Controller Controller 5a 9a 5a...
Page 76: ...72 eFlow rapid 2021 04 3 6 7 60 3 6 7...
Page 77: ...eFlow rapid 2021 04 73 el 8 8 Controller 8 8 6 ml...
Page 78: ...74 eFlow rapid 2021 04...
Page 80: ...76 eFlow rapid 2021 04 Controller Controller 83 eFlow rapid 1 ml 1 ml...
Page 81: ...eFlow rapid 2021 04 77 el 78...
Page 82: ...78 eFlow rapid 2021 04 5 3 5 37 C 5 5 3 easycare 2 easycare...
Page 83: ...eFlow rapid 2021 04 79 el 6...
Page 84: ...80 eFlow rapid 2021 04 5 81...
Page 87: ...eFlow rapid 2021 04 83 el 6 Controller ON OFF 20 ON OFF...
Page 89: ...eFlow rapid 2021 04 85 el easycare easycare easycare easycare...
Page 90: ...86 eFlow rapid 2021 04...
Page 94: ...90 eFlow rapid 2021 04 EN 60601 1 eFlow rapid II BF IEC 60529 IP IP 21...
Page 96: ...92 eFlow rapid 2021 04 10 BF CE 93 42 13 2005...
Page 97: ...eFlow rapid 2021 04 93 el 11 PARI PARI PARI PARI...
Page 140: ...136 eFlow rapid 2021 04 Gakktu r skugga um a allar tengingar s u ttar og a lyfh lfi s l st...
Page 190: ...186 eFlow rapid 2021 04...
Page 250: ...246 eFlow rapid 2021 04...
Page 281: ......