TROUBLESHOOTING
Faults in installation or operation (e.g. neb-
uliser handset not connected, no medica-
tion added, power supply interrupted) are
displayed approx. 5 to 6 times in flashing
mode. The device then switches off. If sev-
eral causes of malfunction exist, they will be
displayed in the following order each time
the device is switched on again. Please fol-
low the step-by-step instructions for cor-
recting such malfunctions (see following
list):
Display indicator
(flashes alternately)
Possible cause
Solution
Battery flat.
Insert new batteries or
operate with the power
adapter connected.
Cable connection to the
control unit or on the
nebuliser handset inter-
rupted.
Check cable
connections.
No medication added.
Add medication.
Inhalation interrupted.
Resume inhalation by
restarting the device
(press ON/OFF button).
Battery charge level
50%.
Have new batteries on
hand or operate with the
power adapter
connected.
The maximum operat-
ing time of 20 min. for
one application has
been exceeded.
Resume inhalation by
restarting the device
(press ON/OFF button).
Summary of Contents for eflowrapid 178G1005
Page 1: ...Gebrauchsanweisung Instructions for Use nebuliser system...
Page 6: ...2 eFlow rapid 03 12...
Page 38: ...34 eFlow rapid 03 12...
Page 70: ...66 eFlow rapid 03 12...
Page 102: ...98 eFlow rapid 03 12...
Page 134: ...130 eFlow rapid 03 12...
Page 166: ...162 eFlow rapid 03 12...
Page 198: ...194 eFlow rapid 03 12...
Page 230: ...226 eFlow rapid 03 12...
Page 262: ...258 eFlow rapid 03 12...
Page 294: ...290 eFlow rapid 03 12...
Page 329: ......