16

      

        

Questions about Lamira?   Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) 

 

 

QUESTION 1: How long should my ARIKAYCE treatment 

take? 

Answer:

 With normal operation and proper cleaning, your 

Handset should deliver 1 vial of ARIKAYCE in about 14 

minutes but could take up to 20 minutes. Your Lamira 

should automatically shut off shortly after your treatment is 

complete or after 20 minutes.  If the Lamira turns off after 

20 minutes, check the Medication Reservoir.  If it is empty, 

you have received your full dose

 

QUESTION 2: How much ARIKAYCE should be left in 

the Medication Reservoir at the end of my 

treatment? 

Answer:

 Only a drop of ARIKAYCE should remain in the 

Medication Reservoir.  If more than a drop remains, start 

the Controller and complete your dose.  Then clean the 

Aerosol Head (Section F).  If after proper cleaning, more 

than a drop remains in the Medication Reservoir, contact 

the Arikares Support Program at 1-833-ARIKARE (1-833-

274-5273). 

QUESTION 3: When I turn on my Nebulizer, nothing 

happens.  There does not seem to be any power. 

Answer:

 If you are using batteries, use the figures in each 

Battery Compartment to check that the batteries are 

inserted correctly.  If the batteries are positioned properly, 

check to see if the light is blinking on the Controller 

signaling low battery power.  If it is, replace the batteries or 

use the A/C Power Supply. 

Answer:

 If you are using the A/C Power Supply, check 

the connection to be sure it is firmly connected to the 

Power Supply Port located underneath the Battery 

Compartment. 

Be sure that the A/C Power Supply is plugged into a 

working wall outlet. 

QUESTION 4:  Sometimes I have trouble removing the 

Medication Cap at the end of my therapy session. 

Answer:

 A vacuum has formed in the Medication 

Reservoir.  Remove the Connection Cord, open the 

Handset and carefully remove the Aerosol Head.  The 

Medication Cap will then be easier to remove. 

QUESTION 5: What if no mist is coming out of your 

Handset? 

Answer

First, check that the Controller has power.  

Secondly, make sure the Connection Cord is correctly 

attached.  Thirdly, check to be sure that your Handset 

has been assembled properly (Section H).  Lastly, check 

to be sure that the ARIKAYCE is in the Medication 

Reservoir.  If the Controller or your Handset still do not 

function properly, contact the Arikares Support Program 

at 1-833-ARIKARE (1-833-274-5273). 

 

 

 

QUESTION 6: What if liquid is leaking from my Handset 

during my treatment? 

Answer

To prevent leaks, be sure (1) the Blue Valve is 

properly attached to the Aerosol Chamber, (2) the Aerosol 

Head is inserted correctly, (3) the Medication Cap is 

screwed on fully, and (4) the Medication Reservoir and 

Aerosol Chamber are closed properly and snapped 

together. 

 

Answer

It is normal to have some liquid collect in the 

Aerosol Chamber.  Try to hold the Handset so that liquid 

does not pour out of the Mouthpiece. 

QUESTION 7: What if my Controller shuts off before my 

treatment begins, or does not restart? 

Answer

:

  Low voltage.  Replace the batteries or use 

the AC adapter.  Press the On/Off button to continue 

your treatment. 

Answer

:

 Your Handset was tilted above 45°.  Hold 

your Handset level and press the On/Off button to 

continue your treatment. 

Answer

:

  ARIKAYCE is cold.  Allow the ARIKAYCE to 

warm to room temperature and then press the On/Off 

button to continue your treatment.  

QUESTION 8: What if my Controller does not shut off at 

the end of my treatment? 

Answer

:  

Your Controller may take up to 60 seconds to 

shut off after you complete your dose.  If you wish to stop 

your Controller earlier, press the On/Off button. 

 

Answer

Disconnect the A/C Power Supply from the 

Controller and remove the batteries.  Then, reinsert the 

Batteries following the procedure in Section D. 

 

Answer

 If this situation continues 3 or 4 times, contact 

the Arikares Support Program at 1-833-ARIKARE (1-

833-274-5273). 

QUESTION 9: How long will a new set of batteries last? 

Answer:

 A new set of batteries should provide 2 hours 

of total use. 

 

Answer:

 Consider using rechargeable batteries. 

QUESTION 10: What if something arrives damaged? 

Answer:

 Contact the Arikares Support Program at 1-833-

ARIKARE (1-833-274-5273). 

 

   

 

 

 

     Frequently Asked Questions 

Reference ID: 4327567

Summary of Contents for Lamira

Page 1: ...Reference ID 4327567...

Page 2: ...Lamira nebulizer Before using your Lamira nebulizer read and understand all the Instructions for Use and save them for future reference If you do not understand any part of these directions contact t...

Page 3: ...ion clean and dry all parts of the Handset after each use Follow the instructions in Section E to maintain and clean the Lamira Cleaning the Handset properly will help prevent the Aerosol Head from cl...

Page 4: ...a Handset to be used for 28 days until the next ARIKAYCE drug box arrives o 1 ARIKAYCE Quick Start Guide o 1 Instructions For Use Insert o 1 Full Prescribing Information Insert Step C2 Check your Lami...

Page 5: ...recommended to remove the batteries to reduce the risk of battery leakage Using the A C power supply The A C Power Supply will automatically adjust to the incoming voltage and will power the Controlle...

Page 6: ...r Mouthpiece Wipe with clean disposable paper towel 1 second per part Wipe to remove residual medicine and then dispose of paper towel in trash with solid waste Rinse Prior to first use and after each...

Page 7: ...ached to the mouthpiece as shown in the picture Your Handset including the Aerosol Head is not sterile Contamination and moisture may cause the growth of bacteria and the Aerosol Head can be affected...

Page 8: ...systems Use only plain dry paper towels or wipes Do not use towels or wipes that have any chemicals added to them such as alcohol lotion or baby wipes Be careful not to harm the parts Do not wipe the...

Page 9: ...e from soap disinfect your Handset Step F10 Disinfect the Handset including the Aerosol Head after cleaning In addition to cleaning ARIKAYCE from your Handset you must also disinfect your Handset to r...

Page 10: ...Section F Step G2 Disinfect your Handset at the end of every day by boiling in distilled water To disinfect with boiling water boil the Handset parts including the Aerosol Head in a clean pot of disti...

Page 11: ...ater and dry them well H 2 Step H3 Open the Handset The Medication Reservoir and the Aerosol Chamber are attached using a hinge First gently pull up on the tab of the Medication Reservoir H 3 This wil...

Page 12: ...e Aerosol Head so that For Amikacin Liposome Inhalation Suspension is facing toward you and is at the top of the Aerosol Head H 5b Squeeze the two flexible plastic tabs together while inserting the me...

Page 13: ...Connection Cord by lining up the bottom of the Connector with the bottom of the Handset H 8a and pushing the Connection Cord upward against the rear underside of the Handset H 8b until you hear the pa...

Page 14: ...me throughout and well mixed I 1 Step I2 Lift orange cap from vial and throw away dispose of the orange cap I 2 Step I3 Grip metal ring on top of the vial Pull it down gently I 3 until 1 side breaks a...

Page 15: ...on the LCD display I 8a You will also hear 1 beep and the status light will turn green The Lamira is now On After treatment begins the treatment screen I 8b will replace the start screen I 8a and aero...

Page 16: ...change from dose to dose and may become longer unless the cleaning and maintenance instructions are followed See Maintaining Your Lamira Step I10 At the end of your treatment the following will happen...

Page 17: ...at the Controller has power Secondly make sure the Connection Cord is correctly attached Thirdly check to be sure that your Handset has been assembled properly Section H Lastly check to be sure that t...

Page 18: ...tart Controller 6 Flashing Lamira beeps 1x then beeps high low with the LED flashing and the LCD display flashing this symbol No ARIKAYCE detected Add ARIKAYCE If you have already added ARIKAYCE gentl...

Page 19: ...el gently tap Handset and restart Controller 16 The Lamira does not stop automatically after all ARIKAYCE is consumed Clean and disinfect the Aerosol Head If condition continues contact the Arikares S...

Page 20: ...imulation of label claimc 52 9 46 4 59 4 MMADd by NGIe m 4 45 4 38 4 52 GSDf 1 59 1 59 1 60 Respiratory Dose by NGI 5 m mg 327 1 316 5 337 6 Respiratory Fraction by NGI 5 m of delivered dose 55 6 53 8...

Reviews: