5
Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273)
Maintaining Your Lamira
E
To reduce the risk of infection, illness, or injury from contamination or improper use, it is important to complete the following 2
steps:
1)
Rinse and clean the Handset including the Aerosol Head before first use and right after each use. (see Section F) Do not
wash the Controller, Connection Cord, or A/C Power Supply. Use clear liquid soap made for washing dishes to clean the Handset.
Do not use liquid dish soaps that are white or antibacterial liquid dish soaps because these may contain additives harmful to the
Aerosol Head.
2)
Disinfect the Handset every day (see Section G).
•
Do not put the Handset or the Aerosol Head in the microwave oven.
•
Do not try to clean the Handset or Aerosol Head in a dishwasher.
•
Do not try to clean the Handset or the Aerosol Head with brushes or abrasives.
Handset Maintenance Summary
(see the next page for complete instructions)
Instruction
When
Parts cleaned
Method
How
Long
Comments
Wipe
After
each use
•
Medication Reservoir and
Aerosol Chamber
•
Mouthpiece
Wipe with clean
disposable
paper towel.
1 second
per part
Wipe to remove residual
medicine and then
dispose of paper towel in
trash with solid waste.
Rinse
Prior to
first use
and after
each use
•
Aerosol Head
•
Medication Cap and Seal
•
Blue Valve
•
Medication Reservoir and
Aerosol Chamber
•
Mouthpiece
Warm running
tap water.
10 seconds Rinse each side of the
Aerosol Head for 10
seconds.
Clean
Prior to
first use
and after
each use
•
Aerosol Head
•
Medication Cap and Seal
•
Blue Valve
•
Medication Reservoir and
Aerosol Chamber
•
Mouthpiece
Soak each
piece in warm
soapy water.
While soaking
swish or shake
each piece.
5 minutes
Use 3 to 5 drops of clear
liquid dish soap in a bowl
with enough warm water
to cover all pieces.
Soak longer if Handset
has dried or if visibly
dirty.
Rinse
Prior to
first use
and after
each use
•
Aerosol Head
•
Medication Cap and Seal
•
Blue Valve
•
Medication Reservoir and
Aerosol Chamber
•
Mouthpiece
Warm running
tap water.
Until soap is
removed.
Check each part and
soak for another 5
minutes if any part looks
dirty.
Disinfect
Prior to
first use
and after
each use
•
Aerosol Head
•
Medication Cap and Seal
•
Blue Valve
•
Medication Reservoir and
Aerosol Chamber
•
Mouthpiece
Boil in distilled
water.
5 minutes
Air-dry in a dust-free
environment.
Caution
Warning
Reference ID: 4327567
Summary of Contents for Lamira
Page 1: ...Reference ID 4327567...