Questions about Lamira?   Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273) 

 

                             Never let the Controller come in contact with 

water or cleaning agents.  If liquid does get into the Controller, 

contact the Arikares Support Program.  

 

 

 

Step F8:

 

Clean all Handset parts by adding a few 

drops of clear liquid dish soap and warm tap 

water to a clean tub or bowl.  Cover the 

Handset parts in the warm soapy water and 

soak for 5 minutes, shaking them periodically 

(F-8).  

 
 
 
 

Step F9:

 

Rinse all parts thoroughly under warm 

running tap water to remove soap (F-9).  

Check each part and soak for another 5 

minutes if any of the parts look dirty.  After all 

parts are cleaned and free from soap, 

disinfect your Handset.  

 
 

Step F10:

 

Disinfect the Handset including the 

Aerosol Head after cleaning.  In addition to 

cleaning ARIKAYCE from your Handset, you 

must also disinfect your Handset to remove 

bacteria and avoid infection.  See Section G 

for instructions on how to disinfect your 

Lamira Handset. 

 

Cleaning your Controller and Connection 

Cord 

 
Step F11:

 

To reduce the risk of electric shock, 

disconnect all connections before 

cleaning.  Switch off the Controller.  

Remove the Connection Cord and A/C 

Power Supply cord from the Controller.   

 

Step F12:

 

Clean the Controller housing and 

Connection Cord as needed with a soft, 

clean, damp cloth.  Do not place the 

Controller unit under water or allow liquid to 

get inside the Controller.  Make sure moisture 

from the cloth does not enter the Controller. 

 

 
 
 

 

  Caution 

F-8

 

F-9

 

     Cleaning Your Lamira (continued) 

Reference ID: 4327567

Summary of Contents for Lamira

Page 1: ...Reference ID 4327567...

Page 2: ...Lamira nebulizer Before using your Lamira nebulizer read and understand all the Instructions for Use and save them for future reference If you do not understand any part of these directions contact t...

Page 3: ...ion clean and dry all parts of the Handset after each use Follow the instructions in Section E to maintain and clean the Lamira Cleaning the Handset properly will help prevent the Aerosol Head from cl...

Page 4: ...a Handset to be used for 28 days until the next ARIKAYCE drug box arrives o 1 ARIKAYCE Quick Start Guide o 1 Instructions For Use Insert o 1 Full Prescribing Information Insert Step C2 Check your Lami...

Page 5: ...recommended to remove the batteries to reduce the risk of battery leakage Using the A C power supply The A C Power Supply will automatically adjust to the incoming voltage and will power the Controlle...

Page 6: ...r Mouthpiece Wipe with clean disposable paper towel 1 second per part Wipe to remove residual medicine and then dispose of paper towel in trash with solid waste Rinse Prior to first use and after each...

Page 7: ...ached to the mouthpiece as shown in the picture Your Handset including the Aerosol Head is not sterile Contamination and moisture may cause the growth of bacteria and the Aerosol Head can be affected...

Page 8: ...systems Use only plain dry paper towels or wipes Do not use towels or wipes that have any chemicals added to them such as alcohol lotion or baby wipes Be careful not to harm the parts Do not wipe the...

Page 9: ...e from soap disinfect your Handset Step F10 Disinfect the Handset including the Aerosol Head after cleaning In addition to cleaning ARIKAYCE from your Handset you must also disinfect your Handset to r...

Page 10: ...Section F Step G2 Disinfect your Handset at the end of every day by boiling in distilled water To disinfect with boiling water boil the Handset parts including the Aerosol Head in a clean pot of disti...

Page 11: ...ater and dry them well H 2 Step H3 Open the Handset The Medication Reservoir and the Aerosol Chamber are attached using a hinge First gently pull up on the tab of the Medication Reservoir H 3 This wil...

Page 12: ...e Aerosol Head so that For Amikacin Liposome Inhalation Suspension is facing toward you and is at the top of the Aerosol Head H 5b Squeeze the two flexible plastic tabs together while inserting the me...

Page 13: ...Connection Cord by lining up the bottom of the Connector with the bottom of the Handset H 8a and pushing the Connection Cord upward against the rear underside of the Handset H 8b until you hear the pa...

Page 14: ...me throughout and well mixed I 1 Step I2 Lift orange cap from vial and throw away dispose of the orange cap I 2 Step I3 Grip metal ring on top of the vial Pull it down gently I 3 until 1 side breaks a...

Page 15: ...on the LCD display I 8a You will also hear 1 beep and the status light will turn green The Lamira is now On After treatment begins the treatment screen I 8b will replace the start screen I 8a and aero...

Page 16: ...change from dose to dose and may become longer unless the cleaning and maintenance instructions are followed See Maintaining Your Lamira Step I10 At the end of your treatment the following will happen...

Page 17: ...at the Controller has power Secondly make sure the Connection Cord is correctly attached Thirdly check to be sure that your Handset has been assembled properly Section H Lastly check to be sure that t...

Page 18: ...tart Controller 6 Flashing Lamira beeps 1x then beeps high low with the LED flashing and the LCD display flashing this symbol No ARIKAYCE detected Add ARIKAYCE If you have already added ARIKAYCE gentl...

Page 19: ...el gently tap Handset and restart Controller 16 The Lamira does not stop automatically after all ARIKAYCE is consumed Clean and disinfect the Aerosol Head If condition continues contact the Arikares S...

Page 20: ...imulation of label claimc 52 9 46 4 59 4 MMADd by NGIe m 4 45 4 38 4 52 GSDf 1 59 1 59 1 60 Respiratory Dose by NGI 5 m mg 327 1 316 5 337 6 Respiratory Fraction by NGI 5 m of delivered dose 55 6 53 8...

Reviews: