CAUTION
Inadequate disinfection encourages the growth of bacteria
and thus increases the
risk of infection
. Adequate cleaning
with disinfection can only be assured if the specified mixing
ratio and application time are adhered to, and if all individual
parts are completely immersed in the solution for the entire
application time. There must not be any air pockets or
bubbles.
• Prepare a 2% solution of
Bomix
®
plus by mixing 10 ml of
the concentrate with 500 ml
tap water.
• Place all the individual parts in
the prepared solution and
leave them to soak for 5 min.
If the application period is exceeded significantly, the
plastic parts may take on the smell of the disinfectant.
• Rinse off all parts thoroughly in running water (residues of
the disinfectant can cause allergic reactions or irritations of
the mucous membrane).
• You can remove excess water more quickly by shaking all
parts.
• Dispose of the used solution (the diluted solution can be got
rid of down the drain).
4.5 Visual inspection
Inspect all product components after each cleaning and disin-
fection. Replace any broken, misshapen or seriously dis-
coloured parts.
– 46 –
Summary of Contents for LC series
Page 1: ...LC Family...
Page 4: ...SONSTIGES 24 Entsorgung 24 Technische Daten 24 Zeichenerkl rung 25 Kontakt 25 4...
Page 26: ...26...
Page 28: ...MISCELLANEOUS 47 Disposal 47 Technical data 47 Explanation of symbols 48 Contact 48 28...
Page 50: ...ANDET 69 Bortskaffelse 69 Tekniske data 69 Tegnforklaring 70 Kontakt 70 50...
Page 96: ...MUUTA 115 K yt st poistaminen 115 Tekniset tiedot 115 Merkkien selitys 116 Yhteystiedot 116 96...
Page 142: ...ALTRO 162 Smaltimento 162 Dati tecnici 162 Spiegazione dei simboli 163 Contatti 163 142...
Page 164: ...164...
Page 210: ...DIVERSOS 229 Elimina o 229 Dados t cnicos 229 Explica o dos s mbolos 230 Contacto 230 210...
Page 232: ...VRIGT 251 Kassering 251 Tekniska data 251 Symbolf rklaring 252 Kontakt 252 232...
Page 253: ......