6.2 Nebuliser data
Operating gases
Air, oxygen
Minimum flow
3.0 l/min.
Corresponding minimum operating pressure
0.5 bar / 50 kPa
Maximum flow
6.0 l/min.
Corresponding maximum operating pressure
2.0 bar / 200 kPa
Minimum fill volume
2 ml
Maximum fill volume
8 ml
Aerosol data according to DIN EN 13544-1 is available on re-
quest from PARI GmbH.
6.3 Classification according to DIN EN 60601-1
Type of electric shock protection
Protection class II
Degree of protection from electric shock of
the part used (nebuliser)
Type BF
Degree of protection against water ingress
in accordance with EN 60529 (IP rating)
No protection
Degree of protection when used in the
presence of flammable mixtures of
anaesthetics with air, with oxygen, or with
nitrous oxide
No protection
Operating mode
Continuous operation
– 79 –